<i>Objective.</i>The introduction of automatic tube current modulation (ATCM) has resulted in complex relationships between scanner parameters, patient body habitus, radiation dose, and image quality. ATCM adjusts tube current based on x-ray attenuation variations in the scan region, and overall patient dose depends on a combination of factors. This work aims to develop mathematical models that predict CT radiation dose and image noise in terms of attenuating diameter and all relevant scanner parameters.<i>Approach.</i>A homogenous phantom, equipped with the features to conduct discrete and continuous adaption tests, was developed to model ATCM in a Philips CT scanner. Scanner parameters were varied based on theoretical dose relationships, and a MATLAB script was developed to extract data from DICOM images. R statistical software was employed for data analysis, plotting, and regression modelling.<i>Main Results.</i>Phantom data provided the following insights: Median tube current decreased by 81% as tube potential varied from 80 kVp to 140 kVp. Doubling the DoseRight Index (DRI) from 12 to 24, at 24 cm diameter, produced a 294% increase in mA and a 46% decrease in noise. Mean mA increased by 53% whilst mean noise increased by 5.7% as helical pitch increased from 0.6 to 0.925. Changing rotation time from 0.33s to 0.75s gave a 56% reduction in mean mA and no change in image noise. Increasing detector collimation (<i>n × T</i>) resulted in higher tube currents and lower output image noise values, as<i>n</i>and<i>T</i>were varied independently. Interpreting these results to apply transformations relevant to each independent variable produced models for tube current and noise with adjusted R-squared values of 0.965 and 0.912, respectively.<i>Significance.</i>The models developed more accurately predict radiation dose and image quality for specific patients and scanner settings. They provide imaging professionals with a practical tool to optimize scan protocols according to patient diameters and clinical objectives.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-10-10
DOI
https://doi.org/10.1088/2057-1976/ad857a
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.18
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of automatic tube current modulation in a CT scanner using a customised homogeneous phantom

Jeyasingam Jeyasugiththan, Aruna Pallewatte, A.M.C. Kumarihami, T Amalaraj et al.
1 citations
Biomedical Physics & Engineering Express
Advanced X-ray and CT Imaging
0.18
article

Evaluation of automatic tube current modulation in a CT scanner using a customised homogeneous phantom

Jeyasingam Jeyasugiththan, Aruna Pallewatte, A.M.C. Kumarihami, T Amalaraj, S Peterson, Urshella Hishaam, Duminda Satharasinghe, Mariana Beligoli Gomes da Costa, Sameera Viswakula
article en
1 citations

Abstract

<i>Objective.</i>The introduction of automatic tube current modulation (ATCM) has resulted in complex relationships between scanner parameters, patient body habitus, radiation dose, and image quality. ATCM adjusts tube current based on x-ray attenuation variations in the scan region, and overall patient dose depends on a combination of factors. This work aims to develop mathematical models that predict CT radiation dose and image noise in terms of attenuating diameter and all relevant scanner parameters.<i>Approach.</i>A homogenous phantom, equipped with the features to conduct discrete and continuous adaption tests, was developed to model ATCM in a Philips CT scanner. Scanner parameters were varied based on theoretical dose relationships, and a MATLAB script was developed to extract data from DICOM images. R statistical software was employed for data analysis, plotting, and regression modelling.<i>Main Results.</i>Phantom data provided the following insights: Median tube current decreased by 81% as tube potential varied from 80 kVp to 140 kVp. Doubling the DoseRight Index (DRI) from 12 to 24, at 24 cm diameter, produced a 294% increase in mA and a 46% decrease in noise. Mean mA increased by 53% whilst mean noise increased by 5.7% as helical pitch increased from 0.6 to 0.925. Changing rotation time from 0.33s to 0.75s gave a 56% reduction in mean mA and no change in image noise. Increasing detector collimation (<i>n × T</i>) resulted in higher tube currents and lower output image noise values, as<i>n</i>and<i>T</i>were varied independently. Interpreting these results to apply transformations relevant to each independent variable produced models for tube current and noise with adjusted R-squared values of 0.965 and 0.912, respectively.<i>Significance.</i>The models developed more accurately predict radiation dose and image quality for specific patients and scanner settings. They provide imaging professionals with a practical tool to optimize scan protocols according to patient diameters and clinical objectives.

Biomedical Physics & Engineering ExpressVol. 10(6)
University of Cape Town (ZA), General Sir John Kotelawala Defence University (LK), National Hospital of Sri Lanka (LK), University of Colombo (LK)
Affordable and clean energy
Openalex Percentile: Top 52%
Advanced X-ray and CT Imaging
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.