Clinical usability and efficacy of a robotic bone fracture reduction system: A pilot animal study

Challenges in minimally invasive surgeries, such as intramedullary nailing for long bone fractures, include radiation overexposure for patients and surgeons, potential malreduction, and physical burden on surgeons in maintaining the reduction status. A robotic bone fracture reduction system was developed in this study to address these problems. The system consists of a hexapod with six degrees of freedom, with a fracture reduction device and a master device. This study aimed to evaluate the novel system in a preclinical setting. The length of the six axes in the system can be adjusted to precisely control the length, angle, and rotation so that no additional traction is required. Fluoroscopic images can be remotely examined to reduce the risk of radiation exposure for surgeons. In this study, alignment accuracy and radiation exposure were measured using 32 bovine bone fracture models, and these surgical outcomes were compared to those of conventional manual surgery to verify the clinical usability and effectiveness of the system. The alignment accuracy was assessed by analyzing length, angulation, and rotation. The four surgeons participating in this study were divided into two groups (expert and novice) according to their clinical experience. All parameters in robotic surgery significantly decreased by approximately 4 mm and 8° on average (p ≤ 0.05) compared to conventional surgery. The mean radiation exposure in robot-assisted surgery was 0.11 mSv, showing a significant decrease compared to conventional surgery (p < 0.05). Reduction accuracy was higher in robotic surgery performed by the novice group than in conventional surgery performed by the expert group; however, standard deviation values were inversed. In conclusion, the bone fracture reduction robot system increased the alignment accuracy through precise control while reducing radiation exposure in surgeons, as the surgery was performed remotely. The use of this system is predicted to improve the accuracy and reproducibility of the surgery and the safety of medical staff.<sup>1</sup>.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2024-10-10
DOI
https://doi.org/10.1016/j.medengphy.2024.104242
Citations
2
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.77

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical usability and efficacy of a robotic bone fracture reduction system: A pilot animal study

Hyunhee Bang, Chang‐Wug Oh, Suk-Joong Lee, Sanghyun Joung et al.
2 citations
Medical Engineering & Physics
Bone fractures and treatments
0.77
article

Clinical usability and efficacy of a robotic bone fracture reduction system: A pilot animal study

Hyunhee Bang, Chang‐Wug Oh, Suk-Joong Lee, Sanghyun Joung, Hyun‐Joo Lee, Joon‐Woo Kim, Il‐Hyung Park
article en
2 citations

Abstract

Challenges in minimally invasive surgeries, such as intramedullary nailing for long bone fractures, include radiation overexposure for patients and surgeons, potential malreduction, and physical burden on surgeons in maintaining the reduction status. A robotic bone fracture reduction system was developed in this study to address these problems. The system consists of a hexapod with six degrees of freedom, with a fracture reduction device and a master device. This study aimed to evaluate the novel system in a preclinical setting. The length of the six axes in the system can be adjusted to precisely control the length, angle, and rotation so that no additional traction is required. Fluoroscopic images can be remotely examined to reduce the risk of radiation exposure for surgeons. In this study, alignment accuracy and radiation exposure were measured using 32 bovine bone fracture models, and these surgical outcomes were compared to those of conventional manual surgery to verify the clinical usability and effectiveness of the system. The alignment accuracy was assessed by analyzing length, angulation, and rotation. The four surgeons participating in this study were divided into two groups (expert and novice) according to their clinical experience. All parameters in robotic surgery significantly decreased by approximately 4 mm and 8° on average (p ≤ 0.05) compared to conventional surgery. The mean radiation exposure in robot-assisted surgery was 0.11 mSv, showing a significant decrease compared to conventional surgery (p < 0.05). Reduction accuracy was higher in robotic surgery performed by the novice group than in conventional surgery performed by the expert group; however, standard deviation values were inversed. In conclusion, the bone fracture reduction robot system increased the alignment accuracy through precise control while reducing radiation exposure in surgeons, as the surgery was performed remotely. The use of this system is predicted to improve the accuracy and reproducibility of the surgery and the safety of medical staff.<sup>1</sup>.

Medical Engineering & PhysicsVol. 133(1)
Kyungpook National University Hospital (KR), Gyeongsang National University Hospital (KR), Kyungpook National University Medical Center (KR), Korea Air Force Academy (KR)
Ministry of Health and Welfare, Korea Health Industry Development Institute
Zero hunger
Openalex Percentile: Top 25%
Bone fractures and treatments
0.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.