Feasibility of shortening scan duration of 18F-FDG myocardial metabolism imaging using a total-body PET/CT scanner
Sufficient image quality for clinical diagnosis could be achieved at G120 for MMI using a total-body PET/CT scanner, while the image quality of G30 was acceptable for myocardial viability analysis.