An exploratory study of the impact of CT slice thickness and inter-rater variability on anatomical accuracy of malunited distal radius models and surgical guides for corrective osteotomy

High-resolution CT images are essential in clinical practice to accurately replicate patient anatomy for 3D virtual surgical planning and designing patient-specific surgical guides. These technologies are commonly used in corrective osteotomy of the distal radius. This study evaluated how the virtual radius models and the surgical guides’ surface that is in contact with the bone vary between experienced raters. Further, the discrepancies from the reference radius of surgical guides and radius models created from CT images with slice thicknesses larger than the reference standard of 0.625mm were assessed. Maximum overlap with radius model was measured for guides, and absolute average distance error was measured for radius models. The agreement between the lower-resolution guides surface and the raters’ guide surface was evaluated. The average inter-rater guide surface overlap was -0.11mm [95% CI: -0.13–0.09]. The surface of surgical guides designed on CT images with a 1mm slice thickness deviated from the reference radius within the inter-rater range (0.03mm). For slice thicknesses of 1.25mm and 1.5mm, the average guide surface overlap was 0.12mm and 0.15mm, respectively. The average inter-rater radius surface variability was 0.03mm [95% CI: 0.025–0.035]. The discrepancy from the reference of all radius models created from CT images with a slice thickness larger than the reference slice thickness was notably larger than the inter-rater variability but, excluding one case, did not exceed 0.2mm. The results suggest that 1mm CT images are suitable for surgical guide design. While 1.25mm slices are commonly used for virtual planning in hand and forearm surgery, slices larger than 1mm may approach the limit of clinical acceptability. Discrepancies in radius models were below 1mm, likely below clinical relevance.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2024-10-10
DOI
https://doi.org/10.1371/journal.pone.0311805
Citations
1
Primary Topic
Orthopedic Surgery and Rehabilitation
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An exploratory study of the impact of CT slice thickness and inter-rater variability on anatomical accuracy of malunited distal radius models and surgical guides for corrective osteotomy

Anders Björkman, Peter Axelsson, Per Fredrikson, Emilia Gryska et al.
1 citations
PLoS ONE
Orthopedic Surgery and Rehabilitation
0.57
article

An exploratory study of the impact of CT slice thickness and inter-rater variability on anatomical accuracy of malunited distal radius models and surgical guides for corrective osteotomy

Anders Björkman, Peter Axelsson, Per Fredrikson, Emilia Gryska, Katleen Libberecht, Charlotte Stor Swinkels
article en
1 citations

Abstract

High-resolution CT images are essential in clinical practice to accurately replicate patient anatomy for 3D virtual surgical planning and designing patient-specific surgical guides. These technologies are commonly used in corrective osteotomy of the distal radius. This study evaluated how the virtual radius models and the surgical guides’ surface that is in contact with the bone vary between experienced raters. Further, the discrepancies from the reference radius of surgical guides and radius models created from CT images with slice thicknesses larger than the reference standard of 0.625mm were assessed. Maximum overlap with radius model was measured for guides, and absolute average distance error was measured for radius models. The agreement between the lower-resolution guides surface and the raters’ guide surface was evaluated. The average inter-rater guide surface overlap was -0.11mm [95% CI: -0.13–0.09]. The surface of surgical guides designed on CT images with a 1mm slice thickness deviated from the reference radius within the inter-rater range (0.03mm). For slice thicknesses of 1.25mm and 1.5mm, the average guide surface overlap was 0.12mm and 0.15mm, respectively. The average inter-rater radius surface variability was 0.03mm [95% CI: 0.025–0.035]. The discrepancy from the reference of all radius models created from CT images with a slice thickness larger than the reference slice thickness was notably larger than the inter-rater variability but, excluding one case, did not exceed 0.2mm. The results suggest that 1mm CT images are suitable for surgical guide design. While 1.25mm slices are commonly used for virtual planning in hand and forearm surgery, slices larger than 1mm may approach the limit of clinical acceptability. Discrepancies in radius models were below 1mm, likely below clinical relevance.

PLoS ONEVol. 19(10)
Sahlgrenska University Hospital (SE), University of Gothenburg (SE)
Sustainable cities and communities
Openalex Percentile: Top 29%
Orthopedic Surgery and Rehabilitation
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.