Small changes in the transducer position cause a systematic change in cardiac output readings: implications for clinical practice

To systematically evaluate the effect of small changes in transducer position on key hemodynamic variables including CO generated by 4th generation FloTrac software. After cardiac surgery, cardiac output, mean arterial pressure, systemic vascular resistance, and stroke volume variation were measured with 4 generation Flotrac software. The transducer position was randomly placed at the midaxillary plane, 4 cm higher than the midaxillary plane or 4 cm lower than the midaxillary plane. Averages of three measurements were used. Data was available from 20 patients. Cardiac output increased from 4.59 L/min (± 0.92) to 4.78 L/min (± 0.99) with the transducer position at the midaxillary plane to 4 cm higher than the midaxillary plane, and cardiac output decreased to 4.43 L/min (± 0.90) with the transducer 4 cm lower than midaxillary plane (P < 0.001). On the relative scale, CO increased 4.1% (95% CI 3.1-5.0) when comparing the higher transducer level with the midaxillary plane position, and CO decreased 3.4% (95% CI 2.4-4.4) when comparing the midaxillary plane position with the lower transducer level, correspondiong to changes in CO of ≈ 1% per 1 cm change in transducer position. Mean arterial pressure and systemic vascular resistance both changed significantly with transducer position (both P < 0.001), whereas no statistically or clinically significant effect was seen on stroke volume variation (P = 0.98). A four-centimeter change in vertical transducer position induced clinically significant changes in cardiac output measurements by 4th generation FloTrac software. Definitions of optimal cardiac output in goal-directed therapy algorithms require meticulous transducer adjustment and can only be used in the reference patient position.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2024-10-10
DOI
https://doi.org/10.1007/s10877-024-01230-6
Citations
3
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
1.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Small changes in the transducer position cause a systematic change in cardiac output readings: implications for clinical practice

Peter Juhl‐Olsen, Simon Tilma Vistisen, Caroline Dinesen
3 citations
Journal of Clinical Monitoring and Computing
Hemodynamic Monitoring and Therapy
1.72
article

Small changes in the transducer position cause a systematic change in cardiac output readings: implications for clinical practice

Peter Juhl‐Olsen, Simon Tilma Vistisen, Caroline Dinesen
article en
3 citations

Abstract

To systematically evaluate the effect of small changes in transducer position on key hemodynamic variables including CO generated by 4th generation FloTrac software. After cardiac surgery, cardiac output, mean arterial pressure, systemic vascular resistance, and stroke volume variation were measured with 4 generation Flotrac software. The transducer position was randomly placed at the midaxillary plane, 4 cm higher than the midaxillary plane or 4 cm lower than the midaxillary plane. Averages of three measurements were used. Data was available from 20 patients. Cardiac output increased from 4.59 L/min (± 0.92) to 4.78 L/min (± 0.99) with the transducer position at the midaxillary plane to 4 cm higher than the midaxillary plane, and cardiac output decreased to 4.43 L/min (± 0.90) with the transducer 4 cm lower than midaxillary plane (P < 0.001). On the relative scale, CO increased 4.1% (95% CI 3.1-5.0) when comparing the higher transducer level with the midaxillary plane position, and CO decreased 3.4% (95% CI 2.4-4.4) when comparing the midaxillary plane position with the lower transducer level, correspondiong to changes in CO of ≈ 1% per 1 cm change in transducer position. Mean arterial pressure and systemic vascular resistance both changed significantly with transducer position (both P < 0.001), whereas no statistically or clinically significant effect was seen on stroke volume variation (P = 0.98). A four-centimeter change in vertical transducer position induced clinically significant changes in cardiac output measurements by 4th generation FloTrac software. Definitions of optimal cardiac output in goal-directed therapy algorithms require meticulous transducer adjustment and can only be used in the reference patient position.

Journal of Clinical Monitoring and ComputingVol. 39(2)
Aarhus University Hospital (DK)
Aarhus Universitetshospital
Quality Education
Openalex Percentile: Top 15%
Hemodynamic Monitoring and Therapy
1.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Small changes in the transducer position cause a systematic change in cardiac output readings: implications for clinical practice — Peter Juhl‐Olsen, Simon Tilma Vistisen, et al. · Journal of Clinical Monitoring and Computing (2024) | TGRS Research Map | TGRS