Contactless femoral implant stability monitoring in cementless total hip arthroplasty, A step towards clinical implementation

The clinical implementation of currently used devices for intraoperative fixation monitoring of femoral implants via vibration-based methods in cementless total hip arthroplasty is challenging, due to practical and regulatory issues. Motivated by the effectiveness of electromagnetic excitation in similar dental applications, this study investigates the use of electromagnetic excitation for femoral implant stability monitoring during cementless total hip arthroplasty. The results obtained from electromagnetic excitation were largely consistent with reference results obtained through impact excitation, with a Pearson Correlation Coefficient of 0.79 in the 0.1-8 kHz frequency band. Moreover, the peak frequencies obtained via the two methods yielded a relative difference of 0.20 ± 0.22 %. Next, the excitation device was successfully utilized in conjunction with a laser vibrometer to monitor the stability of the femoral implant during an in vitro insertion, proving the feasibility of contactless implant stability monitoring. These results indicate the promising potential of this contactless method for clinical implementation.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2024-10-09
DOI
https://doi.org/10.1016/j.medengphy.2024.104243
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.57

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Contactless femoral implant stability monitoring in cementless total hip arthroplasty, A step towards clinical implementation

Michiel Mulier, Leonard Pastrav, George Athanassoulis Makris, Wim Desmet et al.
1 citations
Medical Engineering & Physics
Orthopaedic implants and arthroplasty
0.57
article

Contactless femoral implant stability monitoring in cementless total hip arthroplasty, A step towards clinical implementation

Michiel Mulier, Leonard Pastrav, George Athanassoulis Makris, Wim Desmet, Kathleen Denis, Georges Frederic Vles
article en
1 citations

Abstract

The clinical implementation of currently used devices for intraoperative fixation monitoring of femoral implants via vibration-based methods in cementless total hip arthroplasty is challenging, due to practical and regulatory issues. Motivated by the effectiveness of electromagnetic excitation in similar dental applications, this study investigates the use of electromagnetic excitation for femoral implant stability monitoring during cementless total hip arthroplasty. The results obtained from electromagnetic excitation were largely consistent with reference results obtained through impact excitation, with a Pearson Correlation Coefficient of 0.79 in the 0.1-8 kHz frequency band. Moreover, the peak frequencies obtained via the two methods yielded a relative difference of 0.20 ± 0.22 %. Next, the excitation device was successfully utilized in conjunction with a laser vibrometer to monitor the stability of the femoral implant during an in vitro insertion, proving the feasibility of contactless implant stability monitoring. These results indicate the promising potential of this contactless method for clinical implementation.

Medical Engineering & PhysicsVol. 133(1)
Universitair Ziekenhuis Leuven (BE), KU Leuven (BE)
KU Leuven
Openalex Percentile: Top 29%
Orthopaedic implants and arthroplasty
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.