A comparative study of shade-matching performance using intraoral scanner, spectrophotometer, and visual assessment
This study aimed to explore the clinical applicability of the shade-matching function in intraoral scanners. This study measured the tooth colors of maxillary anterior dentitions of 83 adults using visual matching, a spectrophotometer, and a scanner according to two color systems: VITA Classical (VC) and VITA 3D-Master (V3D). Agreement between each method was assessed by weighted Cohen's kappa coefficient (K<sub>W</sub>, α = 0.05). For V3D, the overall agreement between the scanner and spectrophotometer (K<sub>W</sub> = 0.498) was higher than that between the scanner and visual matching (K<sub>W</sub> = 0.473). Similarly, the agreement between the scanner and spectrophotometer (K<sub>W</sub> = 0.283) was higher than that between a scanner and visual matching (K<sub>W</sub> = 0.140) for VC. Regarding tooth position, the highest agreement between the scanner and spectrophotometer was observed on the right central incisor (K<sub>W</sub> = 0.542) for V3D. Tooth color measurement with a scanner was comparable to that with a spectrophotometer, especially on the central incisors when using the VITA 3D-Master system. A scanner could serve as an alternative to a spectrophotometer for shade selection. However, color matching should still be visually verified.
Authors
- Joo Hyun Lee (ORCID: https://orcid.org/0000-0002-1010-1097)
- Hee‐Kyung Kim (ORCID: https://orcid.org/0000-0002-9079-6086)
Institutions
- Ajou University (KR)
Publication Details
- Journal
- Scientific Reports
- Published
- 2024-10-09
- DOI
- https://doi.org/10.1038/s41598-024-74354-z
- Citations
- 8
- Primary Topic
- Dental materials and restorations
- Type
- article
- Field-Weighted Citation Impact
- 3.22
Funders
- National Research Foundation
- National Research Foundation of Korea
- Ministry of Science and ICT, South Korea