Neonatal Noninvasive Ventilation Nasal Mask Interface Pressure and the Inter-Individual Variation of Mask Placement

<b>Background:</b> The 2014 American Academy of Pediatrics recommendation for CPAP as an alternative to mechanical ventilation for treatment of neonatal respiratory distress prompted a rapid shift to noninvasive respiratory support. Since most patients receive nasal bubble CPAP a concomitant increase in nasal pressure injuries followed. This prospective observational study aimed to develop strategies to reduce nasal mask pressure injury in neonates by (1) quantifying CPAP mask-interface pressure and (2) assessing placement variability. <b>Methods:</b> A micro-electromechanical systems pressure sensor was modified for contact pressure measurements with silicone embedding and calibrated. The CPAP generator and interface components were sized for a 24-week neonatal simulator. Thirteen neonatal ICU staff placed the simulator on CPAP at 6 cm H<sub>2</sub>O and 8 L/min of flow with no humidification. Pressure was measured at 3 locations (the forehead, nasal bridge, and philtrum) in triplicate (3 measurements/site). Descriptive statistics; a location-specific, one-way analysis of variance with a Tukey post hoc test; and a 2-sample paired <i>t</i> test of the means of the first and last triplicate were performed. <b>Results:</b> Pressure ranged from 12-377.3 mm Hg. The mean [SD] interface pressure at the philtrum was significantly higher than both the nasal bridge and the forehead (philtrum 173.9 [101.3], nasal bridge 67.8 [28.9], and forehead 79.0 [36.9], <i>P</i> < .001). CPAP fixation varied, including bonnet placement, trunk angle, mask compression, use of hook and loop extenders, and level of vigorous bubble feedback achieved. <b>Conclusions:</b> This study developed a modified pressure sensor for quantifying the pressure exerted by a nasal mask on facial skin. Maximum pressures were higher than those previously reported. Inter-individual differences were present in both quantitative and qualitative measures of pressure. Reduction of NIV-associated pressure injuries may be achieved through NIV fixation technique training and improved nasal mask stability and size increments.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2024-10-08
DOI
https://doi.org/10.4187/respcare.12217
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Neonatal Noninvasive Ventilation Nasal Mask Interface Pressure and the Inter-Individual Variation of Mask Placement

Eric A. Nauman, Orlando S. Hoilett, Vivek Narendran, Marty O. Visscher et al.
Respiratory Care
Airway Management and Intubation Techniques
article

Neonatal Noninvasive Ventilation Nasal Mask Interface Pressure and the Inter-Individual Variation of Mask Placement

Eric A. Nauman, Orlando S. Hoilett, Vivek Narendran, Marty O. Visscher, Anne D. Zakrajsek, Lissy Kesterson
article en

Abstract

<b>Background:</b> The 2014 American Academy of Pediatrics recommendation for CPAP as an alternative to mechanical ventilation for treatment of neonatal respiratory distress prompted a rapid shift to noninvasive respiratory support. Since most patients receive nasal bubble CPAP a concomitant increase in nasal pressure injuries followed. This prospective observational study aimed to develop strategies to reduce nasal mask pressure injury in neonates by (1) quantifying CPAP mask-interface pressure and (2) assessing placement variability. <b>Methods:</b> A micro-electromechanical systems pressure sensor was modified for contact pressure measurements with silicone embedding and calibrated. The CPAP generator and interface components were sized for a 24-week neonatal simulator. Thirteen neonatal ICU staff placed the simulator on CPAP at 6 cm H<sub>2</sub>O and 8 L/min of flow with no humidification. Pressure was measured at 3 locations (the forehead, nasal bridge, and philtrum) in triplicate (3 measurements/site). Descriptive statistics; a location-specific, one-way analysis of variance with a Tukey post hoc test; and a 2-sample paired <i>t</i> test of the means of the first and last triplicate were performed. <b>Results:</b> Pressure ranged from 12-377.3 mm Hg. The mean [SD] interface pressure at the philtrum was significantly higher than both the nasal bridge and the forehead (philtrum 173.9 [101.3], nasal bridge 67.8 [28.9], and forehead 79.0 [36.9], <i>P</i> < .001). CPAP fixation varied, including bonnet placement, trunk angle, mask compression, use of hook and loop extenders, and level of vigorous bubble feedback achieved. <b>Conclusions:</b> This study developed a modified pressure sensor for quantifying the pressure exerted by a nasal mask on facial skin. Maximum pressures were higher than those previously reported. Inter-individual differences were present in both quantitative and qualitative measures of pressure. Reduction of NIV-associated pressure injuries may be achieved through NIV fixation technique training and improved nasal mask stability and size increments.

Respiratory CareVol. 70(4)
Cincinnati Children's Hospital Medical Center (US), University of Cincinnati (US), University of Cincinnati Medical Center (US)
Purdue University, Purdue Research Foundation, University of Cincinnati
Openalex Percentile: Top 67%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.