Clonality Determination by Detecting Unmodified Monoclonal Serum Free Light Chains Using On-Probe Extraction Coupled with Liquid Chromatography-High-Resolution Mass Spectrometry

Abstract Background Serum free light chains (FLCs) are an essential clinical biomarker for the diagnosis and monitoring of patients with plasma cell neoplasms. The current widely used immunoassay methods quantify total serum FLCs, which include monoclonal FLCs as well as FLCs in the polyclonal background. Patients with chronic diseases, inflammatory disorders, or renal dysfunction can have elevated total FLCs that lead to ambiguous results. These patients may benefit from a direct measurement of monoclonal FLCs. The purpose of this study was to develop a method that couples on-probe extraction (OPEX) with liquid chromatography-high-resolution mass spectrometry (LC-HR-MS), abbreviated to OPEX-MS, to directly determine the clonality of FLCs. Methods OPEX immunocapture was performed using microprobes loaded with anti-kappa or anti-lambda light chain antibodies. Captured proteins were separated by reversed-phase LC and analyzed by HR-MS. Results Four cohorts of samples from unique patients were tested based on immunoassay FLC results. The LC-HR-MS analysis in the OPEX-MS method provides both a unique retention time along with deconvoluted masses of FLC monomers and dimers for each clone. The study found that 16 out of 49 (33%) kappa FLC elevated samples as well as 83 out of 100 (83%) dual kappa and lambda FLC elevated samples did not have monoclonal FLCs, which is consistent with the knowledge that there is often no clonal population in samples with mildly elevated FLC immunoassay results. Conclusions The OPEX-MS method can serve as a complementary approach to directly determine clonality in patients with difficult-to-interpret FLC immunoassay results.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry
Published
2024-10-08
DOI
https://doi.org/10.1093/clinchem/hvae130
Citations
5
Primary Topic
Multiple Myeloma Research and Treatments
Type
article
Field-Weighted Citation Impact
2.27
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clonality Determination by Detecting Unmodified Monoclonal Serum Free Light Chains Using On-Probe Extraction Coupled with Liquid Chromatography-High-Resolution Mass Spectrometry

Priscilla S.-W. Yeung, Samuel Yang, Ruben Yiqi Luo, David Iberri et al.
5 citations
Clinical Chemistry
Multiple Myeloma Research and Treatments
2.27
article

Clonality Determination by Detecting Unmodified Monoclonal Serum Free Light Chains Using On-Probe Extraction Coupled with Liquid Chromatography-High-Resolution Mass Spectrometry

Priscilla S.-W. Yeung, Samuel Yang, Ruben Yiqi Luo, David Iberri, Carolyn V Wong, Run‐Zhang Shi, Yajing Liu, Ashley Ruan, C. R. Kerr
article en
5 citations

Abstract

Abstract Background Serum free light chains (FLCs) are an essential clinical biomarker for the diagnosis and monitoring of patients with plasma cell neoplasms. The current widely used immunoassay methods quantify total serum FLCs, which include monoclonal FLCs as well as FLCs in the polyclonal background. Patients with chronic diseases, inflammatory disorders, or renal dysfunction can have elevated total FLCs that lead to ambiguous results. These patients may benefit from a direct measurement of monoclonal FLCs. The purpose of this study was to develop a method that couples on-probe extraction (OPEX) with liquid chromatography-high-resolution mass spectrometry (LC-HR-MS), abbreviated to OPEX-MS, to directly determine the clonality of FLCs. Methods OPEX immunocapture was performed using microprobes loaded with anti-kappa or anti-lambda light chain antibodies. Captured proteins were separated by reversed-phase LC and analyzed by HR-MS. Results Four cohorts of samples from unique patients were tested based on immunoassay FLC results. The LC-HR-MS analysis in the OPEX-MS method provides both a unique retention time along with deconvoluted masses of FLC monomers and dimers for each clone. The study found that 16 out of 49 (33%) kappa FLC elevated samples as well as 83 out of 100 (83%) dual kappa and lambda FLC elevated samples did not have monoclonal FLCs, which is consistent with the knowledge that there is often no clonal population in samples with mildly elevated FLC immunoassay results. Conclusions The OPEX-MS method can serve as a complementary approach to directly determine clonality in patients with difficult-to-interpret FLC immunoassay results.

Clinical ChemistryVol. 70(12)
Stanford Health Care (US), Stanford Medicine (US), Stanford University (US)
Good health and well-being
Openalex Percentile: Top 11%
Multiple Myeloma Research and Treatments
2.27
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.