Isometric Force–Time Curve Assessment: Accuracy, Precision, and Repeatability of a Mobile Application and Portable and Lightweight Device

Strength assessment is one of the main fields in sports performance, physical rehabilitation, physical activity, and health. We aimed to compare maximal voluntary isometric contractions (MVICs) and paired voluntary isometric contractions (VICs) of knee extensors between an isokinetic dynamometer (BIODEX) and a portable and lightweight device (DINABANG). From 19 volunteers (age: 28.7 ± 7 years; body mass: 72 ± 10 kg; and height: 173 ± 7 cm) we obtained 114 paired MVIC measures and, from the force–time curves of these repetitions, 22,507 paired VIC measures of knee extensors. We observed “excellent” repeatability for MVICs (ICC:1.00; p < 0.001) between BIODEX (247 ± 79.5 Nm) and DINABANG (247 ± 74.8 Nm), with “trivial” effect (mean difference: 0.12 Nm (0.02%); 95%CI: −0.13 to 0.23 Nm; p = 0.606; d = 0.048). Bland–Altman plots revealed high accuracy for MVIC (bias: 0.12 Nm) and consistent distribution (precision) inside the limits of agreement (−4.81 to 5.06 Nm) and respective 95%CI. “Excellent” repeatability was also observed for VICs (ICC:1.00; p < 0.001) between BIODEX (219 ± 84.1 Nm) and DINABANG (218 ± 84.0 Nm), with “trivial” effect (0.24 Nm (0.11%); 0.08 to 0.11 Nm; p < 0.001; d = 0.100). Bland–Altman plots revealed high accuracy for VICs (bias: 0.24 Nm) and consistent distribution (precision) inside the limits of agreement (−4.5 to 4.9 Nm) and respective 95%CI. DINABANG is accurate, precise, and reliable in torque measurement.

Authors

Institutions

Publication Details

Journal
Sports
Published
2024-10-07
DOI
https://doi.org/10.3390/sports12100268
Citations
3
Primary Topic
Sports Performance and Training
Type
article
Field-Weighted Citation Impact
1.43
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Isometric Force–Time Curve Assessment: Accuracy, Precision, and Repeatability of a Mobile Application and Portable and Lightweight Device

Rodrigo Zacca, Franco Símini, Alfredo Bravo‐Sànchez, Leonardo Alexandre Peyré‐Tartaruga et al.
3 citations
Sports
Sports Performance and Training
1.43
article

Isometric Force–Time Curve Assessment: Accuracy, Precision, and Repeatability of a Mobile Application and Portable and Lightweight Device

Rodrigo Zacca, Franco Símini, Alfredo Bravo‐Sànchez, Leonardo Alexandre Peyré‐Tartaruga, Darío Santos
article en
3 citations

Abstract

Strength assessment is one of the main fields in sports performance, physical rehabilitation, physical activity, and health. We aimed to compare maximal voluntary isometric contractions (MVICs) and paired voluntary isometric contractions (VICs) of knee extensors between an isokinetic dynamometer (BIODEX) and a portable and lightweight device (DINABANG). From 19 volunteers (age: 28.7 ± 7 years; body mass: 72 ± 10 kg; and height: 173 ± 7 cm) we obtained 114 paired MVIC measures and, from the force–time curves of these repetitions, 22,507 paired VIC measures of knee extensors. We observed “excellent” repeatability for MVICs (ICC:1.00; p < 0.001) between BIODEX (247 ± 79.5 Nm) and DINABANG (247 ± 74.8 Nm), with “trivial” effect (mean difference: 0.12 Nm (0.02%); 95%CI: −0.13 to 0.23 Nm; p = 0.606; d = 0.048). Bland–Altman plots revealed high accuracy for MVIC (bias: 0.12 Nm) and consistent distribution (precision) inside the limits of agreement (−4.81 to 5.06 Nm) and respective 95%CI. “Excellent” repeatability was also observed for VICs (ICC:1.00; p < 0.001) between BIODEX (219 ± 84.1 Nm) and DINABANG (218 ± 84.0 Nm), with “trivial” effect (0.24 Nm (0.11%); 0.08 to 0.11 Nm; p < 0.001; d = 0.100). Bland–Altman plots revealed high accuracy for VICs (bias: 0.24 Nm) and consistent distribution (precision) inside the limits of agreement (−4.5 to 4.9 Nm) and respective 95%CI. DINABANG is accurate, precise, and reliable in torque measurement.

SportsVol. 12(10)
Universidade Federal do Rio Grande do Sul (BR), Universidad de la República de Uruguay (UY), Universidade do Porto (PT), Universidad Francisco de Vitoria (ES), University of Pavia (IT), University of Castilla-La Mancha (ES)
Openalex Percentile: Top 20%
Sports Performance and Training
1.43
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.