A Low-Cost, Portable, Multi-Cancer Screening Device Based on a Ratio Fluorometry and Signal Correlation Technique

The autofluorescence of erythrocyte porphyrins has emerged as a potential method for multi-cancer early detection (MCED). With this method's dependence on research-grade spectrofluorometers, significant improvements in instrumentation are necessary to translate its potential into clinical practice, as with any promising medical technology. To fill this gap, in this paper, we present an automated ratio porphyrin analyzer for cancer screening (ARPA-CS), a low-cost, portable, and automated instrument for MCED via the ratio fluorometry of porphyrins. The ARPA-CS aims to facilitate cancer screening in an inexpensive, rapid, non-invasive, and reasonably accurate manner for use in primary clinics or at point of care. To accomplish this, the ARPA-CS uses an ultraviolet-excited optical apparatus for ratio fluorometry that features two photodetectors for detection at 590 and 630 nm. Additionally, it incorporates a synchronous detector for the precision measurement of signals based on the Walsh-ordered Walsh-Hadamard transform (WHT)<sub>w</sub> and circular shift. To estimate its single-photodetector capability, we established a linear calibration curve for the ARBA-CS exceeding four orders of magnitude with a linearity of up to 0.992 and a low detection limit of 0.296 µg/mL for riboflavin. The ARPA-CS also exhibited excellent repeatability (0.21%) and stability (0.60%). Moreover, the ratio fluorometry of three serially diluted standard solutions of riboflavin yielded a ratio of 0.4, which agrees with that expected based on the known emission spectra of riboflavin. Additionally, the ratio fluorometry of the porphyrin solution yielded a ratio of 49.82, which was ascribed to the predominant concentration of protoporphyrin IX in the brown eggshells, as confirmed in several studies. This study validates this instrument for the ratio fluorometry of porphyrins as a biomarker for MCED. Nevertheless, large and well-designed clinical trials are necessary to further elaborate more on this matter.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2024-10-07
DOI
https://doi.org/10.3390/bios14100482
Citations
2
Primary Topic
Photodynamic Therapy Research Studies
Type
article
Field-Weighted Citation Impact
0.84
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Low-Cost, Portable, Multi-Cancer Screening Device Based on a Ratio Fluorometry and Signal Correlation Technique

Abdulaziz S. Alghamdi, Rabah W. Aldhaheri
2 citations
Biosensors
Photodynamic Therapy Research Studies
0.84
article

A Low-Cost, Portable, Multi-Cancer Screening Device Based on a Ratio Fluorometry and Signal Correlation Technique

Abdulaziz S. Alghamdi, Rabah W. Aldhaheri
article en
2 citations

Abstract

The autofluorescence of erythrocyte porphyrins has emerged as a potential method for multi-cancer early detection (MCED). With this method's dependence on research-grade spectrofluorometers, significant improvements in instrumentation are necessary to translate its potential into clinical practice, as with any promising medical technology. To fill this gap, in this paper, we present an automated ratio porphyrin analyzer for cancer screening (ARPA-CS), a low-cost, portable, and automated instrument for MCED via the ratio fluorometry of porphyrins. The ARPA-CS aims to facilitate cancer screening in an inexpensive, rapid, non-invasive, and reasonably accurate manner for use in primary clinics or at point of care. To accomplish this, the ARPA-CS uses an ultraviolet-excited optical apparatus for ratio fluorometry that features two photodetectors for detection at 590 and 630 nm. Additionally, it incorporates a synchronous detector for the precision measurement of signals based on the Walsh-ordered Walsh-Hadamard transform (WHT)<sub>w</sub> and circular shift. To estimate its single-photodetector capability, we established a linear calibration curve for the ARBA-CS exceeding four orders of magnitude with a linearity of up to 0.992 and a low detection limit of 0.296 µg/mL for riboflavin. The ARPA-CS also exhibited excellent repeatability (0.21%) and stability (0.60%). Moreover, the ratio fluorometry of three serially diluted standard solutions of riboflavin yielded a ratio of 0.4, which agrees with that expected based on the known emission spectra of riboflavin. Additionally, the ratio fluorometry of the porphyrin solution yielded a ratio of 49.82, which was ascribed to the predominant concentration of protoporphyrin IX in the brown eggshells, as confirmed in several studies. This study validates this instrument for the ratio fluorometry of porphyrins as a biomarker for MCED. Nevertheless, large and well-designed clinical trials are necessary to further elaborate more on this matter.

BiosensorsVol. 14(10)
King Abdulaziz University (SA)
Good health and well-being
Openalex Percentile: Top 24%
Photodynamic Therapy Research Studies
0.84
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.