Long‐term performance monitoring of a‐Si 1200 electronic portal imaging device for dosimetric applications
Techniques to monitor and assess the long-term performance of the a-Si 1200 EPID as a dosimeter were developed and implemented using two TrueBeam linacs. The long-term variation of the EPID response was within clinical tolerance indicated in AAPM TG-142 report, and the detector was shown to be stable and reproducible for routine clinical dosimetry.