Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy

From the viewpoints of the advantage depths (ADs), peak tumor dose and skin dose, we evaluated the effect on the dose distribution of neutron beam properties, namely the ratio between thermal and epithermal neutron fluxes (thermal/epithermal ratio), fast neutron component and γ-ray component. Several parameter surveys were conducted with respect to the beam properties of neutron sources for boron neutron capture therapy assuming boronophenylalanine as the boron agent using our dose calculation tool, called SiDE. The ADs decreased by 3% at a thermal/epithermal ratio of 20-30% compared with the current recommendation of 5%. The skin dose increased with the increasing thermal/epithermal ratio, reaching a restricted value of 14 Gyeq at a thermal/epithermal ratio of 48%. The fast neutron component was modified using two different models, namely the 'linear model', in which the fast neutron intensity decreases log-linearly with the increasing neutron energy, and the 'moderator thickness (MT) model', in which the fast neutron component is varied by adjusting the MT in a virtual beam shaping assembly. Although a higher fast neutron component indicated a higher skin dose, the increment was <10% at a fast neutron component of <1 × 10-12 Gy cm2 for both models. Furthermore, in the MT model, the epithermal neutron intensity at a fast neutron component of 6.8 × 10-13 Gy cm2 was 41% higher compared with that of 2 × 10-13 Gy cm2. The γ-ray component also caused no significant disadvantages up to several times larger compared with the current recommendation.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Research
Published
2024-10-04
DOI
https://doi.org/10.1093/jrr/rrae076
Citations
2
Primary Topic
Boron Compounds in Chemistry
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy

Yuki Tanagami, Yoshinori Sakurai, Akira Uritani, Akihisa Ishikawa et al.
2 citations
Journal of Radiation Research
Boron Compounds in Chemistry
0.81
article

Effect of neutron beam properties on dose distributions in a water phantom for boron neutron capture therapy

Yuki Tanagami, Yoshinori Sakurai, Akira Uritani, Akihisa Ishikawa, Hiroaki Kumada, Yoshiaki Kiyanagi, Sachiko Yoshihashi, Hiroki Tanaka, Kenichi Watanabe, Satoshi Nakamura
article en
2 citations

Abstract

From the viewpoints of the advantage depths (ADs), peak tumor dose and skin dose, we evaluated the effect on the dose distribution of neutron beam properties, namely the ratio between thermal and epithermal neutron fluxes (thermal/epithermal ratio), fast neutron component and γ-ray component. Several parameter surveys were conducted with respect to the beam properties of neutron sources for boron neutron capture therapy assuming boronophenylalanine as the boron agent using our dose calculation tool, called SiDE. The ADs decreased by 3% at a thermal/epithermal ratio of 20-30% compared with the current recommendation of 5%. The skin dose increased with the increasing thermal/epithermal ratio, reaching a restricted value of 14 Gyeq at a thermal/epithermal ratio of 48%. The fast neutron component was modified using two different models, namely the 'linear model', in which the fast neutron intensity decreases log-linearly with the increasing neutron energy, and the 'moderator thickness (MT) model', in which the fast neutron component is varied by adjusting the MT in a virtual beam shaping assembly. Although a higher fast neutron component indicated a higher skin dose, the increment was <10% at a fast neutron component of <1 × 10-12 Gy cm2 for both models. Furthermore, in the MT model, the epithermal neutron intensity at a fast neutron component of 6.8 × 10-13 Gy cm2 was 41% higher compared with that of 2 × 10-13 Gy cm2. The γ-ray component also caused no significant disadvantages up to several times larger compared with the current recommendation.

Journal of Radiation ResearchVol. 65(6)
Japan Atomic Energy Agency (JP), Kyushu University (JP), Osaka University of Health and Sport Sciences (JP), University of Tsukuba (JP), Hokkaido University (JP), Kyoto University (JP), Japan Proton Accelerator Research Complex (JP), National Cancer Center (US), Research Organization for Information Science and Technology (JP), Nagoya University (JP)
Clean water and sanitation
Openalex Percentile: Top 25%
Boron Compounds in Chemistry
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.