Influence of Tibial Component Design Features and Interference Fit on Implant-Bone Micromotion in Total Ankle Replacement: A Finite Element Study
This study presents new insights into the effect of TAR primary fixation features on implant-bone micromotion. Although design features heavily influenced implant stability in the model, their influence was greatly diminished when interference fit was introduced. The results of this study show the relative importance of design features and interference fit in the predicted initial stability of uncemented TAR, potentially a key factor in implant survivorship.
Authors
- Donald D. Anderson (ORCID: https://orcid.org/0000-0002-1640-6107)
- Joshua E. Johnson (ORCID: https://orcid.org/0000-0002-8049-4105)
- Gabriel Clarke (ORCID: https://orcid.org/0000-0003-0583-4808)
- César de César Netto (ORCID: https://orcid.org/0000-0001-6037-0685)
Institutions
- University of Iowa (US)
- Duke University Hospital (US)
Publication Details
- Journal
- Foot & Ankle International
- Published
- 2024-10-06
- DOI
- https://doi.org/10.1177/10711007241281294
- Citations
- 4
- Primary Topic
- Foot and Ankle Surgery
- Type
- article
- Field-Weighted Citation Impact
- 1.91