7007 Analytical & Clinical Performance of a Novel Sensitive Rapid Bioassay for Thyrotropin Receptor Antibodies

Abstract Disclosure: P.D. Olivo: Consulting Fee; Self; QuidelOrtho, CA USA. H. Kim: Employee; Self; QuidelOrtho. L. Miao: Employee; Self; QuidelOrtho. J. Houtz: Employee; Self; QuidelOrtho. M. Luffy: None. J. Wolf: None. G.J. Kahaly: Grant Recipient; Self; QuidelOrtho, USA. Background: The role of stimulatory anti-TSH-R-Ab in the pathogenesis of autoimmune hyperthyroidism is well established. A novel and rapid cell-based bioassay, referred to as Turbo TSITM, for measurement of thyroid-stimulating autoantibodies (TSAb) was recently developed. An assessment of the analytical & clinical performance of this bioassay is described herein. Methods: Turbo TSI greatly reduces the complexity and time required to measure TSAb. It avoids the need for cell culture, has no wash or lysis steps, and has a total turn-around-time of less than 90 minutes. Thawed cells from Turbo TSI kits were treated with different concentrations of a World Health Organization (WHO) international standard (IS) TSI or TSAb-positive serum. TSAb was measured as a function of luciferase activity measured as relative light units (RLU) and converted into international units per liter (IU/L). Two users at two sites performed analytical performance studies on numerous samples, over multiple days. Results: The limit of blank, limit of detection and limit of quantitation were determined to be 0.007 IU/L, 0.014 IU/L, and 0.021 IU/L, respectively. Receiver operator characteristics (ROC) analysis determined the cut-off to be 0.0241 IU/L with an area under the curve of 0.984. The linear range was shown to be from 0.015 to 11.958 IU/L. The intra-laboratory precision was ≤15 percent CV. The overall reproducibility of the assay was ≤20 percent CV for five concentrations (0.06 to 5.16 IU/L). Interference and cross reactivity studies with a variety of substances showed that the assay was robust. Turbo TSI demonstrated 95.2% (95% CI 83.3-98.1) sensitivity and 94.8% (95% CI 90.9-97.1) specificity when compared with an FDA-cleared bioassay (Thyretain® TSI) using serum from 295 patients with autoimmune thyroid disease (AITD). When testing 172 unselected, consecutive TSAb+ AITD patients, a US based, prospective, multicenter, blinded study showed a 98% concordance between Turbo TSI and Thyretain® TSI. When comparing Turbo TSI, Thyretain TSI and a TSH-R-Ab binding immunoassay (Cobas) in 670 unselected, consecutive, treated and untreated AITD subjects, sensitivity was 83%, 79% and 68%, respectively (Turbo vs. Cobas, P<0.001). Furthermore, testing 84 patients with naïve, untreated Graves’ hyperthyroidism, detection of TSH-R-Ab was 100%, 98% and 92%, respectively (Turbo TSI vs binding assay, P<0.01). Finally, 95% vs. 73% of 332 patients with Graves’ orbitopathy were TSH-R-Ab+ with Turbo TSI vs. binding assay (P<0.001). Conclusions: The Turbo TSI bioassay exhibits excellent analytical & clinical performance, a high level of reproducibility, and compares well with the FDA-cleared Thyretain® TSI. Presentation: 6/1/2024

Authors

Institutions

Publication Details

Journal
Journal of the Endocrine Society
Published
2024-10-01
DOI
https://doi.org/10.1210/jendso/bvae163.1898
Primary Topic
Monoclonal and Polyclonal Antibodies Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

7007 Analytical & Clinical Performance of a Novel Sensitive Rapid Bioassay for Thyrotropin Receptor Antibodies

Paul D. Olivo, Jeffrey A. Houtz, George J. Kahaly, Miao Liu et al.
Journal of the Endocrine Society
Monoclonal and Polyclonal Antibodies Research
article

7007 Analytical & Clinical Performance of a Novel Sensitive Rapid Bioassay for Thyrotropin Receptor Antibodies

Paul D. Olivo, Jeffrey A. Houtz, George J. Kahaly, Miao Liu, Hyun Kim, Maximilian Luffy, Jean‐Eric Wolf
article en

Abstract

Abstract Disclosure: P.D. Olivo: Consulting Fee; Self; QuidelOrtho, CA USA. H. Kim: Employee; Self; QuidelOrtho. L. Miao: Employee; Self; QuidelOrtho. J. Houtz: Employee; Self; QuidelOrtho. M. Luffy: None. J. Wolf: None. G.J. Kahaly: Grant Recipient; Self; QuidelOrtho, USA. Background: The role of stimulatory anti-TSH-R-Ab in the pathogenesis of autoimmune hyperthyroidism is well established. A novel and rapid cell-based bioassay, referred to as Turbo TSITM, for measurement of thyroid-stimulating autoantibodies (TSAb) was recently developed. An assessment of the analytical & clinical performance of this bioassay is described herein. Methods: Turbo TSI greatly reduces the complexity and time required to measure TSAb. It avoids the need for cell culture, has no wash or lysis steps, and has a total turn-around-time of less than 90 minutes. Thawed cells from Turbo TSI kits were treated with different concentrations of a World Health Organization (WHO) international standard (IS) TSI or TSAb-positive serum. TSAb was measured as a function of luciferase activity measured as relative light units (RLU) and converted into international units per liter (IU/L). Two users at two sites performed analytical performance studies on numerous samples, over multiple days. Results: The limit of blank, limit of detection and limit of quantitation were determined to be 0.007 IU/L, 0.014 IU/L, and 0.021 IU/L, respectively. Receiver operator characteristics (ROC) analysis determined the cut-off to be 0.0241 IU/L with an area under the curve of 0.984. The linear range was shown to be from 0.015 to 11.958 IU/L. The intra-laboratory precision was ≤15 percent CV. The overall reproducibility of the assay was ≤20 percent CV for five concentrations (0.06 to 5.16 IU/L). Interference and cross reactivity studies with a variety of substances showed that the assay was robust. Turbo TSI demonstrated 95.2% (95% CI 83.3-98.1) sensitivity and 94.8% (95% CI 90.9-97.1) specificity when compared with an FDA-cleared bioassay (Thyretain® TSI) using serum from 295 patients with autoimmune thyroid disease (AITD). When testing 172 unselected, consecutive TSAb+ AITD patients, a US based, prospective, multicenter, blinded study showed a 98% concordance between Turbo TSI and Thyretain® TSI. When comparing Turbo TSI, Thyretain TSI and a TSH-R-Ab binding immunoassay (Cobas) in 670 unselected, consecutive, treated and untreated AITD subjects, sensitivity was 83%, 79% and 68%, respectively (Turbo vs. Cobas, P<0.001). Furthermore, testing 84 patients with naïve, untreated Graves’ hyperthyroidism, detection of TSH-R-Ab was 100%, 98% and 92%, respectively (Turbo TSI vs binding assay, P<0.01). Finally, 95% vs. 73% of 332 patients with Graves’ orbitopathy were TSH-R-Ab+ with Turbo TSI vs. binding assay (P<0.001). Conclusions: The Turbo TSI bioassay exhibits excellent analytical & clinical performance, a high level of reproducibility, and compares well with the FDA-cleared Thyretain® TSI. Presentation: 6/1/2024

Journal of the Endocrine SocietyVol. 8(Supplement_1)
Johannes Gutenberg University Mainz (DE), Washington University in St. Louis (US), University Medical Center of the Johannes Gutenberg University Mainz (DE), Quidel Corporation (United States) (US)
Università di Pisa
Partnerships for the goals
Openalex Percentile: Top 71%
Monoclonal and Polyclonal Antibodies Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.