Diagnostic Accuracy of a Novel Wireless, Bluetooth-Enabled Simultaneous 12-Lead Portable ECG Device Compared with Standard ECG in Preoperative Patient Population

Export Background and Objectives: Significant proportion of patients undergoing various surgeries have abnormal ECG findings and it is known to have prognostic value. Recently, portable ECG with Bluetooth technology with smartphone have been gaining popularity. One such novel device, Wellnest ® is a Bluetooth enabled portable ECG machine, can capture simultaneous 12 lead ECG on smartphone application. This is probably only such device available device in India. The objective of this study was to test diagnostic accuracy of Wellnest 12L ECG device compared to standard 12 lead ECG. Methods and Material: This was a Prospective, observational, diagnostic accuracy study conducted at a tertiary care level hospital. During, pre-anaesthesia check-up visit of 109 patients undergoing cardiac and non-cardiac surgeries, whenever indicated, a preoperative 12 lead ECG was done. The Wellnest 12 lead ECG was done within 24 hours of standard ECG. Matching was done between Minnesota codes assigned to ECG findings of Wellnest and standard 12 lead ECG. Also, grading of Wellnest ECG quality was done as excellent, good, bad and poor. Results: In all, 216 various Minnesota codes were assigned to 109 patients undergoing various surgeries. The primary endpoint of Wellnest ECG in comparison to standard ECG were- sensitivity of 96.51% (CI 90.14- 99.27%); specificity of 86.96% (CI 66.41- 97.22%); positive likelihood ratio of 7.4 (CI 2.57- 21.27) and negative likelihood ratio of 0.04 (CI 0.01- 0.12). Performance of Wellnest ECG as per individual Minnesota Codes compared to standard ECG was also acceptable. Wellnest ECG was graded as excellent in 55(50%) patients, good in 51(47%) patients and only 3 (3%) patients were graded as bad. Conclusions: Wellnest ECG had satisfactory performance in comparison with standard 12 lead ECG in preoperative patient population. It is easy to use, portable and can give simultaneous 12 leads ECG and its use in preoperative population can be recommended.

Authors

Institutions

Publication Details

Journal
Annals of Cardiac Anaesthesia
Published
2024-10-01
DOI
https://doi.org/10.4103/aca.aca_79_24
Citations
1
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic Accuracy of a Novel Wireless, Bluetooth-Enabled Simultaneous 12-Lead Portable ECG Device Compared with Standard ECG in Preoperative Patient Population

Unmesh Takalkar, Devendra Borgaonkar, Pooja Joshi, Manisha Takalkar et al.
1 citations
Annals of Cardiac Anaesthesia
ECG Monitoring and Analysis
0.40
article

Diagnostic Accuracy of a Novel Wireless, Bluetooth-Enabled Simultaneous 12-Lead Portable ECG Device Compared with Standard ECG in Preoperative Patient Population

Unmesh Takalkar, Devendra Borgaonkar, Pooja Joshi, Manisha Takalkar, Pradnya Shinde, Diksha Wavhal, Sujitkumar Khade, Reshma Bhosale, Antony George, Deepak Borde
article en
1 citations

Abstract

Export Background and Objectives: Significant proportion of patients undergoing various surgeries have abnormal ECG findings and it is known to have prognostic value. Recently, portable ECG with Bluetooth technology with smartphone have been gaining popularity. One such novel device, Wellnest ® is a Bluetooth enabled portable ECG machine, can capture simultaneous 12 lead ECG on smartphone application. This is probably only such device available device in India. The objective of this study was to test diagnostic accuracy of Wellnest 12L ECG device compared to standard 12 lead ECG. Methods and Material: This was a Prospective, observational, diagnostic accuracy study conducted at a tertiary care level hospital. During, pre-anaesthesia check-up visit of 109 patients undergoing cardiac and non-cardiac surgeries, whenever indicated, a preoperative 12 lead ECG was done. The Wellnest 12 lead ECG was done within 24 hours of standard ECG. Matching was done between Minnesota codes assigned to ECG findings of Wellnest and standard 12 lead ECG. Also, grading of Wellnest ECG quality was done as excellent, good, bad and poor. Results: In all, 216 various Minnesota codes were assigned to 109 patients undergoing various surgeries. The primary endpoint of Wellnest ECG in comparison to standard ECG were- sensitivity of 96.51% (CI 90.14- 99.27%); specificity of 86.96% (CI 66.41- 97.22%); positive likelihood ratio of 7.4 (CI 2.57- 21.27) and negative likelihood ratio of 0.04 (CI 0.01- 0.12). Performance of Wellnest ECG as per individual Minnesota Codes compared to standard ECG was also acceptable. Wellnest ECG was graded as excellent in 55(50%) patients, good in 51(47%) patients and only 3 (3%) patients were graded as bad. Conclusions: Wellnest ECG had satisfactory performance in comparison with standard 12 lead ECG in preoperative patient population. It is easy to use, portable and can give simultaneous 12 leads ECG and its use in preoperative population can be recommended.

Annals of Cardiac AnaesthesiaVol. 27(4)
Lourdes Hospital (IN), Seth Nandlal Dhoot Hospital (IN)
Openalex Percentile: Top 32%
ECG Monitoring and Analysis
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.