Post-occlusion surge parameters during ex vivo phacoemulsification with a new method of adaptive infusion control

A new method of adaptive control of infusion during phacoemulsification (PE) based on the Optimed Profi (Optimedservice) surgical system was developed, allowing calculation of the predicted volume of post-occlusion surge (POS) by monitoring aspiration and infusion flow rates. Purpose . To compare POS amplitude, predicted and actual volumes of POS during experimental ex vivo PE on cadaveric porcine eyes, between Optimed Profi and Centurion Vision surgical systems. Material and methods . For two experimental PE series of 10 operations on porcine eyes, the mean POS amplitude (mm Hg) was measured using a pressure sensor in the anterior eye chamber. Predicted and actual POS volumes were assessed using a non-contact aspiration-line flowmeter sensor by varying aspiration flow rate depending on the phaco needle patency. The parameters were compared between the series using the Student t-test. Results . The mean POS amplitude during PE by Optimed Profi and Centurion Vision System was 12.10 ± 0.21 and 13.3 ± 0.3 mm Hg, respectively (p < 0.001). The mean value of the predicted POS volume during PE by Optimed Profi and Centurion Centurion Vision System was 136.80 ± 9.59 and 146.00 ± 9.46 μm, respectively (p > 0.01). The mean value of the actual POS volume during PE by Optimed Profi and Centurion Vision System was 135.30 ± 4.97 and 158.50 ± 8.63 μm, respectively (p < 0.01). Conclusion . The lower values of POS amplitude and predicted and actual volumes of POS in the series using the new method of adaptive infusion control may indicate a better hydrodynamic stability in experimental PE.

Authors

Institutions

Publication Details

Journal
Russian Ophthalmological Journal
Published
2024-10-02
DOI
https://doi.org/10.21516/2072-0076-2024-17-3-79-84
Citations
2
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
1.08
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Post-occlusion surge parameters during ex vivo phacoemulsification with a new method of adaptive infusion control

Т. И. Дибаев, T. N. Ismagilov, Б. М. Азнабаев, Т. Р. Мухамадеев
2 citations
Russian Ophthalmological Journal
Intraocular Surgery and Lenses
1.08
article

Post-occlusion surge parameters during ex vivo phacoemulsification with a new method of adaptive infusion control

Т. И. Дибаев, T. N. Ismagilov, Б. М. Азнабаев, Т. Р. Мухамадеев
article en
2 citations

Abstract

A new method of adaptive control of infusion during phacoemulsification (PE) based on the Optimed Profi (Optimedservice) surgical system was developed, allowing calculation of the predicted volume of post-occlusion surge (POS) by monitoring aspiration and infusion flow rates. Purpose . To compare POS amplitude, predicted and actual volumes of POS during experimental ex vivo PE on cadaveric porcine eyes, between Optimed Profi and Centurion Vision surgical systems. Material and methods . For two experimental PE series of 10 operations on porcine eyes, the mean POS amplitude (mm Hg) was measured using a pressure sensor in the anterior eye chamber. Predicted and actual POS volumes were assessed using a non-contact aspiration-line flowmeter sensor by varying aspiration flow rate depending on the phaco needle patency. The parameters were compared between the series using the Student t-test. Results . The mean POS amplitude during PE by Optimed Profi and Centurion Vision System was 12.10 ± 0.21 and 13.3 ± 0.3 mm Hg, respectively (p < 0.001). The mean value of the predicted POS volume during PE by Optimed Profi and Centurion Centurion Vision System was 136.80 ± 9.59 and 146.00 ± 9.46 μm, respectively (p > 0.01). The mean value of the actual POS volume during PE by Optimed Profi and Centurion Vision System was 135.30 ± 4.97 and 158.50 ± 8.63 μm, respectively (p < 0.01). Conclusion . The lower values of POS amplitude and predicted and actual volumes of POS in the series using the new method of adaptive infusion control may indicate a better hydrodynamic stability in experimental PE.

Russian Ophthalmological JournalVol. 17(3)
Bashkir State Medical University (RU)
Openalex Percentile: Top 22%
Intraocular Surgery and Lenses
1.08
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.