Post-irradiation darkening model for EBT-3 films characterized using a single lot calibration approach

<i>Objective.</i>In this study, we present a model to correct the progressive post-irradiation darkening of EBT3 films. The model allows for a clinical use of EBT3 using application and calibration films scanned with different post-irradiation times.<i>Approach.</i>The model is a post-irradiation time- and dose-dependent power-law function, projecting the scanned transmittance of application films to the transmittance matching the same post-irradiation time of calibration films. The model was characterized for two EBT3 production lots within the dose range 0.1-12.8 Gy. A first characterization was performed utilizing calibration films scanned repeatedly for 54 d post-irradiation (lot 1), while a fast re-characterization of a second lot used three post-irradiation times (lot 2). For a long-term validation of the model, 16 film strips were irradiated at 2 Gy on different time points starting from the day of film calibration up to 43 d afterwards (lot 1). For the multiple-dose validation of the model, 8 strips were irradiated with dose levels ranging 0-12 Gy deposited 25 d after the calibration (lot 2). As a proof of principle, the model was applied to four clinical patient-specific quality assurance film measurements with prescribed dose/fraction ranging 2.66 Gy-8 Gy.<i>Main results</i>. The post-irradiation transmittance decreased for higher doses up to -2.5% at 12.8 Gy, and 54 d post-irradiation. With a lot-specific model correction, the mean dose accuracy of validation strips that ranged from initial -3.4% (triple-channel) and -9.90% (blue-channel) reduced to within 3.0% (all colour channels) for doses above 1 Gy. The median dose difference with the planned dose improved from -3.5% to -1.1%, and the 3%/2 mm local gamma ranged from (48.5-92.5)% to (81.2-99.2)%.<i>Significance.</i>A film darkening model corrects the transmittance of EBT3 films and increases the flexibility of existing dosimetry protocols. The correction ensures dose accuracies within 3%.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-10-02
DOI
https://doi.org/10.1088/1361-6560/ad8295
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Post-irradiation darkening model for EBT-3 films characterized using a single lot calibration approach

Robin De Roover, Laurence Delombaerde, Wouter Crijns, Marco Caprioli
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Post-irradiation darkening model for EBT-3 films characterized using a single lot calibration approach

Robin De Roover, Laurence Delombaerde, Wouter Crijns, Marco Caprioli
article en

Abstract

<i>Objective.</i>In this study, we present a model to correct the progressive post-irradiation darkening of EBT3 films. The model allows for a clinical use of EBT3 using application and calibration films scanned with different post-irradiation times.<i>Approach.</i>The model is a post-irradiation time- and dose-dependent power-law function, projecting the scanned transmittance of application films to the transmittance matching the same post-irradiation time of calibration films. The model was characterized for two EBT3 production lots within the dose range 0.1-12.8 Gy. A first characterization was performed utilizing calibration films scanned repeatedly for 54 d post-irradiation (lot 1), while a fast re-characterization of a second lot used three post-irradiation times (lot 2). For a long-term validation of the model, 16 film strips were irradiated at 2 Gy on different time points starting from the day of film calibration up to 43 d afterwards (lot 1). For the multiple-dose validation of the model, 8 strips were irradiated with dose levels ranging 0-12 Gy deposited 25 d after the calibration (lot 2). As a proof of principle, the model was applied to four clinical patient-specific quality assurance film measurements with prescribed dose/fraction ranging 2.66 Gy-8 Gy.<i>Main results</i>. The post-irradiation transmittance decreased for higher doses up to -2.5% at 12.8 Gy, and 54 d post-irradiation. With a lot-specific model correction, the mean dose accuracy of validation strips that ranged from initial -3.4% (triple-channel) and -9.90% (blue-channel) reduced to within 3.0% (all colour channels) for doses above 1 Gy. The median dose difference with the planned dose improved from -3.5% to -1.1%, and the 3%/2 mm local gamma ranged from (48.5-92.5)% to (81.2-99.2)%.<i>Significance.</i>A film darkening model corrects the transmittance of EBT3 films and increases the flexibility of existing dosimetry protocols. The correction ensures dose accuracies within 3%.

Physics in Medicine and BiologyVol. 69(20)
KU Leuven (BE)
Agentschap Innoveren en Ondernemen
Openalex Percentile: Top 75%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.