Infrared emissivity measurement methods considering target reflective characteristics

Emissivity measurements are of great significance for infrared thermal radiation and infrared remote sensing. However, traditional methods often face challenges such as difficulties in non-contact measurement, small measurement areas, and unsuitability for non-Lambertian surfaces. To address these issues, we propose the reflective distribution model integral method (RDMIM). This method is based on a reflective distribution model using the scattering-reflective deviation angle (SRDA). By regressing and integrating the object’s reflective distribution model, it achieves accurate non-contact measurement of non-Lambertian surfaces under normal temperature conditions. Additionally, the measurement scheme has been further optimized to improve measurement efficiency while ensuring the accuracy of the model regression. Finally, the proposed RDMIM method has been validated through experimental measurements. The results have shown that this method has advantages in non-contact and large-area measurements. Moreover, the systematic error is smaller when the reflective characteristics of the reference body and the target are relatively similar.

Authors

Institutions

Publication Details

Journal
Applied Optics
Published
2024-10-02
DOI
https://doi.org/10.1364/ao.535336
Citations
4
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
5.20
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Infrared emissivity measurement methods considering target reflective characteristics

Xiumei Fan, Puyousen Zhang, Yin Liu, Simin Ma et al.
4 citations
Applied Optics
Calibration and Measurement Techniques
5.20
article

Infrared emissivity measurement methods considering target reflective characteristics

Xiumei Fan, Puyousen Zhang, Yin Liu, Simin Ma, Yiwen Li, Yao Li, li weiqin
article en
4 citations

Abstract

Emissivity measurements are of great significance for infrared thermal radiation and infrared remote sensing. However, traditional methods often face challenges such as difficulties in non-contact measurement, small measurement areas, and unsuitability for non-Lambertian surfaces. To address these issues, we propose the reflective distribution model integral method (RDMIM). This method is based on a reflective distribution model using the scattering-reflective deviation angle (SRDA). By regressing and integrating the object’s reflective distribution model, it achieves accurate non-contact measurement of non-Lambertian surfaces under normal temperature conditions. Additionally, the measurement scheme has been further optimized to improve measurement efficiency while ensuring the accuracy of the model regression. Finally, the proposed RDMIM method has been validated through experimental measurements. The results have shown that this method has advantages in non-contact and large-area measurements. Moreover, the systematic error is smaller when the reflective characteristics of the reference body and the target are relatively similar.

Applied OpticsVol. 63(31)
Air Force Engineering University (CN), Xi'an University of Technology (CN)
Climate action
Openalex Percentile: Top 5%
Calibration and Measurement Techniques
5.20
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Infrared emissivity measurement methods considering target reflective characteristics — Xiumei Fan, Puyousen Zhang, et al. · Applied Optics (2024) | TGRS Research Map | TGRS