Investigating adsorption and removal of divalent Ca2+ and Pb2+ ions from aqueous solutions by gamma-irradiation using quartz tuning fork (QTF) sensor technique

In this study, gold-coated quartz tuning forks (QTFs) sensing devices functionalized with self-assembled monolayers (SAM) of a lower-rim functionalized calix[4]arene methoxy ester were used for the detection of divalent Ca 2+ and Pb 2+ ions in aqueous solutions by utilizing adsorption behavior and the radiative effect. The gold-coated QTF functionalized calix[4]arene methoxy ester sensing device was tested by measuring the respective frequency shifts obtained using small (60 µL) samples of aqueous PbCl 2 at two different concentrations (10 −6 and 10 −4 M). For 10 −4 M solutions of PbCl 2 , results showed that the resonance frequency shift Δf = 317 Hz, from 32,867 Hz (f Calix ) to 32,550 Hz (f Calix⊃Pb 2+ ) due to the absorption of lead (Pb 2+ ) ions (10 -4 M) by calixl[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. The most significant frequency changes were observed at a concentration of 10 −6 M CaCl 2 , where CaCl 2 exhibited the biggest change of 356 Hz, from 32,893 Hz (f Calix ) to 32,537 Hz (f Calix⊃Ca 2+ ), compared to 317 Hz for PbCl 2 (10 −4 M). The limit of detection was 100 femtomolar (fM) for CaCl 2 and 245 fM for PbCl 2 . After that, we irradiated the receptor molecules which was holding Pb 2+ ions in the complex ([calix ⊃ Pb 2+ ]) on the QTF sensing layer with a radiation dose ranging from 7.5 to 50 µGy of gamma rays from the Cesium-137 source for 30 min. Interestingly, it was observed that the resonance frequency shift (Δf = 54 Hz) back to 32,604 Hz from 32,550 Hz (f (Calix⊃Pb 2+ ) ), which strongly suggests that the Pb 2+ ion removed from ([calix ⊃ Pb 2+ ]) complex on the QTF sensing layer due to gamma radiation dose. To follow up on the radiation effect of the ([calix ⊃ Pb 2+ ]) on the QTF sensing layer, we stopped the gamma radiation source and kept it for an additional 10 min to see if there was any resonance frequency. It was noticed that an additional resonance frequency shifted (Δf = 33 Hz) back to 32,637 Hz from 32,604 Hz after stopping the gamma radiation source for 10 min. We assume that the complex ([calix ⊃ Pb 2+ ]) absorbs the gamma radiation and continues the removal of Pb 2+ ions from the complex on the sensing layer. A similar phenomenon was also observed for the absorption of lead (Pb 2+ ) ions (10 -6 M) by calix[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. The resonance frequency shift Δf = 142 Hz, from 32,706 Hz (f Calix ) to 32,564 Hz (f Calix⊃Pb 2+ ) due to the absorption of lead (Pb 2+ ) ions (10 -6 M) by calix[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. Subsequent exposure to gamma radiation, with doses ranging from 7.5 to 50 µGy from a Cesium-137 source over 30 min, prompted a resonance frequency shift (Δf = 37 Hz) back to 32,606 Hz from 32,564 Hz (f (Calix⊃Pb 2+ ) ). This shift strongly suggests the removal of Pb 2+ ions from the ([calix ⊃ Pb 2+ ]) complex on the QTF sensing layer due to gamma radiation dose. X-ray photoelectron spectroscopy (XPS) analysis confirmed the chemical adsorption of Pb 2+ ions onto the gold-coated QTFs functionalized calix[4]arene adsorbent sensing layer surface. Therefore, the technology underlying the calix[4]arene-functionalized sensing device holds promise for diverse industrial applications, supporting potential advancements in water pollution mitigation through the proposed adsorption and irradiation mechanism.

Authors

Institutions

Publication Details

Journal
Journal of Saudi Chemical Society
Published
2024-10-02
DOI
https://doi.org/10.1016/j.jscs.2024.101947
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigating adsorption and removal of divalent Ca2+ and Pb2+ ions from aqueous solutions by gamma-irradiation using quartz tuning fork (QTF) sensor technique

Shofiur Rahman, Abdullah N. Alodhayb, Khalid E. Alzahrani, Nadyah Alanazi et al.
Journal of Saudi Chemical Society
Advanced Chemical Sensor Technologies
article

Investigating adsorption and removal of divalent Ca2+ and Pb2+ ions from aqueous solutions by gamma-irradiation using quartz tuning fork (QTF) sensor technique

Shofiur Rahman, Abdullah N. Alodhayb, Khalid E. Alzahrani, Nadyah Alanazi, Nahed Alarifi, Mahmoud Al‐Gawati, Reem Alanazi
article en

Abstract

In this study, gold-coated quartz tuning forks (QTFs) sensing devices functionalized with self-assembled monolayers (SAM) of a lower-rim functionalized calix[4]arene methoxy ester were used for the detection of divalent Ca 2+ and Pb 2+ ions in aqueous solutions by utilizing adsorption behavior and the radiative effect. The gold-coated QTF functionalized calix[4]arene methoxy ester sensing device was tested by measuring the respective frequency shifts obtained using small (60 µL) samples of aqueous PbCl 2 at two different concentrations (10 −6 and 10 −4 M). For 10 −4 M solutions of PbCl 2 , results showed that the resonance frequency shift Δf = 317 Hz, from 32,867 Hz (f Calix ) to 32,550 Hz (f Calix⊃Pb 2+ ) due to the absorption of lead (Pb 2+ ) ions (10 -4 M) by calixl[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. The most significant frequency changes were observed at a concentration of 10 −6 M CaCl 2 , where CaCl 2 exhibited the biggest change of 356 Hz, from 32,893 Hz (f Calix ) to 32,537 Hz (f Calix⊃Ca 2+ ), compared to 317 Hz for PbCl 2 (10 −4 M). The limit of detection was 100 femtomolar (fM) for CaCl 2 and 245 fM for PbCl 2 . After that, we irradiated the receptor molecules which was holding Pb 2+ ions in the complex ([calix ⊃ Pb 2+ ]) on the QTF sensing layer with a radiation dose ranging from 7.5 to 50 µGy of gamma rays from the Cesium-137 source for 30 min. Interestingly, it was observed that the resonance frequency shift (Δf = 54 Hz) back to 32,604 Hz from 32,550 Hz (f (Calix⊃Pb 2+ ) ), which strongly suggests that the Pb 2+ ion removed from ([calix ⊃ Pb 2+ ]) complex on the QTF sensing layer due to gamma radiation dose. To follow up on the radiation effect of the ([calix ⊃ Pb 2+ ]) on the QTF sensing layer, we stopped the gamma radiation source and kept it for an additional 10 min to see if there was any resonance frequency. It was noticed that an additional resonance frequency shifted (Δf = 33 Hz) back to 32,637 Hz from 32,604 Hz after stopping the gamma radiation source for 10 min. We assume that the complex ([calix ⊃ Pb 2+ ]) absorbs the gamma radiation and continues the removal of Pb 2+ ions from the complex on the sensing layer. A similar phenomenon was also observed for the absorption of lead (Pb 2+ ) ions (10 -6 M) by calix[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. The resonance frequency shift Δf = 142 Hz, from 32,706 Hz (f Calix ) to 32,564 Hz (f Calix⊃Pb 2+ ) due to the absorption of lead (Pb 2+ ) ions (10 -6 M) by calix[4]arene methoxy ester receptor molecules on the QTF sensing layer from the aqueous solution to forming the ([Calix ⊃ Pb 2+ ]) complex. Subsequent exposure to gamma radiation, with doses ranging from 7.5 to 50 µGy from a Cesium-137 source over 30 min, prompted a resonance frequency shift (Δf = 37 Hz) back to 32,606 Hz from 32,564 Hz (f (Calix⊃Pb 2+ ) ). This shift strongly suggests the removal of Pb 2+ ions from the ([calix ⊃ Pb 2+ ]) complex on the QTF sensing layer due to gamma radiation dose. X-ray photoelectron spectroscopy (XPS) analysis confirmed the chemical adsorption of Pb 2+ ions onto the gold-coated QTFs functionalized calix[4]arene adsorbent sensing layer surface. Therefore, the technology underlying the calix[4]arene-functionalized sensing device holds promise for diverse industrial applications, supporting potential advancements in water pollution mitigation through the proposed adsorption and irradiation mechanism.

Journal of Saudi Chemical SocietyVol. 28(6)
King Saud University (SA)
King Saud University
Clean water and sanitation
Openalex Percentile: Top 84%
Advanced Chemical Sensor Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.