Enhancing auxetic gradient structures for hip joint implants to optimize stress shielding reduction

Abstract This study investigates the design and optimization of a porous hip implant to mitigate stress shielding. Initially, the focus was on determining the elastic modulus of a three-dimensional auxetic structure, primarily in the y-direction. Various methods—numerical, analytical, and experimental—were used to assess the elastic properties. Additive manufacturing was employed to create samples, which were then tested for their elastic properties through compression testing. The results revealed a strong correlation between the elastic modulus values obtained from simulations and experimental tests in the y-direction. To further enhance the implant’s performance and reduce stress shielding at the implant-bone interface, a gradient structure was introduced. This gradient design progressively increases the elastic modulus away from the bone contact surfaces, aligning closely with the bone’s modulus at the interface. The elastic modulus of this gradient structure was computed using Abaqus software and validated through analytical methods in MATLAB, with a minimal 4.8% difference between the two approaches, demonstrating high agreement. The application of a genetic algorithm enabled the creation of a porous hip implant tailored to minimize stress shielding throughout its structure. This innovative approach, integrating numerical, analytical, and experimental techniques with gradient structures, holds promise for improving hip implant performance and enhancing patient outcomes by reducing stress-shielding complications.

Authors

Institutions

Publication Details

Journal
Physica Scripta
Published
2024-09-30
DOI
https://doi.org/10.1088/1402-4896/ad818e
Citations
8
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
4.58
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing auxetic gradient structures for hip joint implants to optimize stress shielding reduction

Hamed Barati, Hussain Gharehbaghi, Sahel Mohammadi Ghalehney, Mohammad Sadeghi
8 citations
Physica Scripta
Orthopaedic implants and arthroplasty
4.58
article

Enhancing auxetic gradient structures for hip joint implants to optimize stress shielding reduction

Hamed Barati, Hussain Gharehbaghi, Sahel Mohammadi Ghalehney, Mohammad Sadeghi
article en
8 citations

Abstract

Abstract This study investigates the design and optimization of a porous hip implant to mitigate stress shielding. Initially, the focus was on determining the elastic modulus of a three-dimensional auxetic structure, primarily in the y-direction. Various methods—numerical, analytical, and experimental—were used to assess the elastic properties. Additive manufacturing was employed to create samples, which were then tested for their elastic properties through compression testing. The results revealed a strong correlation between the elastic modulus values obtained from simulations and experimental tests in the y-direction. To further enhance the implant’s performance and reduce stress shielding at the implant-bone interface, a gradient structure was introduced. This gradient design progressively increases the elastic modulus away from the bone contact surfaces, aligning closely with the bone’s modulus at the interface. The elastic modulus of this gradient structure was computed using Abaqus software and validated through analytical methods in MATLAB, with a minimal 4.8% difference between the two approaches, demonstrating high agreement. The application of a genetic algorithm enabled the creation of a porous hip implant tailored to minimize stress shielding throughout its structure. This innovative approach, integrating numerical, analytical, and experimental techniques with gradient structures, holds promise for improving hip implant performance and enhancing patient outcomes by reducing stress-shielding complications.

Physica ScriptaVol. 99(11)
University of Nebraska at Omaha (US), Tarbiat Modares University (IR), Amirkabir University of Technology (IR)
Openalex Percentile: Top 5%
Orthopaedic implants and arthroplasty
4.58
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.