Evaluation of the Colistin HiMIC Plate Kit for Colistin Susceptibility Testing of <i>Klebsiella pneumoniae</i>

Colistin HiMIC Plate Kit (HiMedia Laboratories), a new commercial broth microdilution (BMD) test for colistin susceptibility testing was evaluated. BMD according to ISO standard 20776-1 (2019) with two-fold dilutions from 128 to 0.125 mg/L was used as a reference method. The colistin reference MICs (minimal inhibitory concentration) ranged from 0,25 to 128 mg/L with 15 (20.5%; 15/73) isolates having colistin reference MICs close to the current EUCAST breakpoint (MICs of 2, 4, and 8 mg/L). The study assessed the compliance of a commercial kit with the CLSI criteria, including categorical agreement (CA) and essential agreement (EA ≥90%), very major error (VME rate) <3%, and major error (ME) rate <3%. On 73 carbapenemase-producing <i>Klebsiella pneumoniae</i> isolates Colistin HiMIC<sup>TM</sup> Plate Kit showed CA and EA of 100% (73/73; 95% CI: 0.97-1.00) and 82.2% (60/73; 95% CI: 0.72-0.90), respectively. No ME (false-resistant results) and VME (false-susceptible results) were detected. Kit showed acceptable CA, ME, and VME error parameters, whereas the EA did not meet the ≥90% threshold. Laboratories must check for possible limitations of commercial kits before they can be used for colistin susceptibility testing.

Authors

Institutions

Publication Details

Journal
Microbial Drug Resistance
Published
2024-09-30
DOI
https://doi.org/10.1089/mdr.2024.0141
Primary Topic
Antibiotic Resistance in Bacteria
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Colistin HiMIC Plate Kit for Colistin Susceptibility Testing of Klebsiella pneumoniae

Tomislav Kuliš, Ana Budimir, Ivana Mareković, Lucija Kanižaj
Microbial Drug Resistance
Antibiotic Resistance in Bacteria
article

Evaluation of the Colistin HiMIC Plate Kit for Colistin Susceptibility Testing of Klebsiella pneumoniae

Tomislav Kuliš, Ana Budimir, Ivana Mareković, Lucija Kanižaj
article en

Abstract

Colistin HiMIC Plate Kit (HiMedia Laboratories), a new commercial broth microdilution (BMD) test for colistin susceptibility testing was evaluated. BMD according to ISO standard 20776-1 (2019) with two-fold dilutions from 128 to 0.125 mg/L was used as a reference method. The colistin reference MICs (minimal inhibitory concentration) ranged from 0,25 to 128 mg/L with 15 (20.5%; 15/73) isolates having colistin reference MICs close to the current EUCAST breakpoint (MICs of 2, 4, and 8 mg/L). The study assessed the compliance of a commercial kit with the CLSI criteria, including categorical agreement (CA) and essential agreement (EA ≥90%), very major error (VME rate) <3%, and major error (ME) rate <3%. On 73 carbapenemase-producing <i>Klebsiella pneumoniae</i> isolates Colistin HiMIC<sup>TM</sup> Plate Kit showed CA and EA of 100% (73/73; 95% CI: 0.97-1.00) and 82.2% (60/73; 95% CI: 0.72-0.90), respectively. No ME (false-resistant results) and VME (false-susceptible results) were detected. Kit showed acceptable CA, ME, and VME error parameters, whereas the EA did not meet the ≥90% threshold. Laboratories must check for possible limitations of commercial kits before they can be used for colistin susceptibility testing.

Microbial Drug ResistanceVol. 30(11)
University Hospital Centre Zagreb (HR), University of Zagreb (HR)
Zero hunger
Openalex Percentile: Top 90%
Antibiotic Resistance in Bacteria
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.