Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments

<b>Objectives:</b> The objective of the present study was to evaluate the cyclic fatigue strength of clockwise cutting rotary endodontic instruments when subjected to two different kinematics: continuous clockwise rotation and clockwise reciprocation movement under optimum torque reverse (OTR) motion. <b>Methods:</b> New ProTaper Next X1 (n = 20) and X2 (n = 20) instruments were randomly divided into two subgroups (n = 10) based on kinematics (continuous rotation or OTR). The specimens were tested using a custom-made device with a non-tapered stainless-steel artificial canal measuring 19 mm in length, featuring a 6 mm radius and an 86-degree curvature. All instruments were tested with a lubricant at room temperature until a fracture occurred. The time to fracture and the length of the separated fragment were recorded. Subsequently, the fractured instruments were inspected under a scanning electron microscope for signs of cyclic fatigue failure, plastic deformation, and/or crack propagation. The subgroup comparisons for time to fracture and instrument length were performed using the independent samples <i>t</i>-test, with the level of statistical significance set at 0.05. <b>Results:</b> When using OTR movement, the ProTaper Next X1 increased the time to fracture from 52.9 to 125.8 s (<i>p</i> < 0.001), while the ProTaper Next X2 increased from 45.4 to 66.0 s (<i>p</i> < 0.001). No subgroup exhibited plastic deformations, but both showed dimpling marks indicative of cyclic fatigue as the primary mode of failure. Additionally, OTR movement resulted in more metal alloy microcracks. <b>Conclusions:</b> The use of OTR motion extended the lifespan of the tested instruments and resulted in a higher number of metal microcracks. This suggests that OTR motion helped to distribute the mechanical stress more evenly across the instrument, thereby relieving localized tension. As a result, it delayed the formation of a single catastrophic crack, enhancing the overall performance of the instruments during the experimental procedures.

Authors

Institutions

Publication Details

Journal
Dentistry Journal
Published
2024-09-30
DOI
https://doi.org/10.3390/dj12100317
Citations
4
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
1.97
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments

Marco A. Versiani, Duarte Marques, Emmanuel João Nogueira Leal Silva, Jorge N. R. Martins
4 citations
Dentistry Journal
Endodontics and Root Canal Treatments
1.97
article

Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments

Marco A. Versiani, Duarte Marques, Emmanuel João Nogueira Leal Silva, Jorge N. R. Martins
article en
4 citations

Abstract

<b>Objectives:</b> The objective of the present study was to evaluate the cyclic fatigue strength of clockwise cutting rotary endodontic instruments when subjected to two different kinematics: continuous clockwise rotation and clockwise reciprocation movement under optimum torque reverse (OTR) motion. <b>Methods:</b> New ProTaper Next X1 (n = 20) and X2 (n = 20) instruments were randomly divided into two subgroups (n = 10) based on kinematics (continuous rotation or OTR). The specimens were tested using a custom-made device with a non-tapered stainless-steel artificial canal measuring 19 mm in length, featuring a 6 mm radius and an 86-degree curvature. All instruments were tested with a lubricant at room temperature until a fracture occurred. The time to fracture and the length of the separated fragment were recorded. Subsequently, the fractured instruments were inspected under a scanning electron microscope for signs of cyclic fatigue failure, plastic deformation, and/or crack propagation. The subgroup comparisons for time to fracture and instrument length were performed using the independent samples <i>t</i>-test, with the level of statistical significance set at 0.05. <b>Results:</b> When using OTR movement, the ProTaper Next X1 increased the time to fracture from 52.9 to 125.8 s (<i>p</i> < 0.001), while the ProTaper Next X2 increased from 45.4 to 66.0 s (<i>p</i> < 0.001). No subgroup exhibited plastic deformations, but both showed dimpling marks indicative of cyclic fatigue as the primary mode of failure. Additionally, OTR movement resulted in more metal alloy microcracks. <b>Conclusions:</b> The use of OTR motion extended the lifespan of the tested instruments and resulted in a higher number of metal microcracks. This suggests that OTR motion helped to distribute the mechanical stress more evenly across the instrument, thereby relieving localized tension. As a result, it delayed the formation of a single catastrophic crack, enhancing the overall performance of the instruments during the experimental procedures.

Dentistry JournalVol. 12(10)
University of Lisbon (PT), Universidade Federal Fluminense (BR), Universidade Unigranrio (BR), Universidade do Estado do Rio de Janeiro (BR), Hospital Central da Polícia Militar (BR)
Openalex Percentile: Top 14%
Endodontics and Root Canal Treatments
1.97
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.