A note on electrometer intercomparison: a routine quality assurance check

Background: This study evaluates the accuracy of the Max-4000 and Model 206 electrometers in measuring ionization readings during radiation therapy dosimetry. The goal is to assess their performance and reliability to determine if they can provide interchangeable and reliable measurements. These comparisons are crucial to ensure accurate radiation dose delivery to patients, improving treatment outcomes and patient safety in clinical settings. Methods: Measurements were conducted using a Varian TrueBeam Edge at different photon energies (6 MV, 6 FFF, 10 MV, and 10 FFF) and field sizes ranging from 3.0 to 30.0 cm. The setup included a solid water phantom with an ionization chamber positioned at 5 cm depth and a 95 cm source-to-surface distance. Both electrometers, ADCL calibrated, were compared for accuracy and precision using the Bland-Altman method, regression analysis, and correlation analysis. Results: The stability of individual charge readings was within ±0.005% for all energies, field sizes, and electrometers. The Bland-Altman analysis showed a bias of -0.0282 [0.0158 to -0.0405, 95% confidence interval (CI)] with a correlation coefficient (R²) of 0.999 (p<0.001) for 6 MV, -0.0345 [0.0283 to -0.0408, 95% CI] with an R² of 0.999 (p<0.001) for 6 FFF, -0.0373 [0.0305 to -0.0441, 95% CI] with an R² of 0.999 (p<0.001) for 10 MV, and -0.00454 [0.00360 to -0.0127, 95% CI] with an R² of 0.999 (p<0.001) for 10 FFF. Conclusions: this study establishes a framework for comparing two ADCL electrometers using Bland-Altman analysis, linear regression, and mean absolute percentage error (MAPE).

Authors

Institutions

Publication Details

Journal
International Journal of Research in Medical Sciences
Published
2024-09-30
DOI
https://doi.org/10.18203/2320-6012.ijrms20242911
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A note on electrometer intercomparison: a routine quality assurance check

Patrice L. Kibisu, Aime M Gloi
International Journal of Research in Medical Sciences
Advanced Chemical Sensor Technologies
article

A note on electrometer intercomparison: a routine quality assurance check

Patrice L. Kibisu, Aime M Gloi
article en

Abstract

Background: This study evaluates the accuracy of the Max-4000 and Model 206 electrometers in measuring ionization readings during radiation therapy dosimetry. The goal is to assess their performance and reliability to determine if they can provide interchangeable and reliable measurements. These comparisons are crucial to ensure accurate radiation dose delivery to patients, improving treatment outcomes and patient safety in clinical settings. Methods: Measurements were conducted using a Varian TrueBeam Edge at different photon energies (6 MV, 6 FFF, 10 MV, and 10 FFF) and field sizes ranging from 3.0 to 30.0 cm. The setup included a solid water phantom with an ionization chamber positioned at 5 cm depth and a 95 cm source-to-surface distance. Both electrometers, ADCL calibrated, were compared for accuracy and precision using the Bland-Altman method, regression analysis, and correlation analysis. Results: The stability of individual charge readings was within ±0.005% for all energies, field sizes, and electrometers. The Bland-Altman analysis showed a bias of -0.0282 [0.0158 to -0.0405, 95% confidence interval (CI)] with a correlation coefficient (R²) of 0.999 (p<0.001) for 6 MV, -0.0345 [0.0283 to -0.0408, 95% CI] with an R² of 0.999 (p<0.001) for 6 FFF, -0.0373 [0.0305 to -0.0441, 95% CI] with an R² of 0.999 (p<0.001) for 10 MV, and -0.00454 [0.00360 to -0.0127, 95% CI] with an R² of 0.999 (p<0.001) for 10 FFF. Conclusions: this study establishes a framework for comparing two ADCL electrometers using Bland-Altman analysis, linear regression, and mean absolute percentage error (MAPE).

International Journal of Research in Medical SciencesVol. 12(10)
Sutter Health (US)
Openalex Percentile: Top 85%
Advanced Chemical Sensor Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.