Finite element analysis and optimization studies on tibia implant of SS 316L steel and Ti6Al4V alloy

Tibial fractures account for approximately 15% of all fractures, typically resulting from high-energy trauma. A critical surgical approach to treat these fractures involves the fixation of the tibia using a plate with minimally invasive osteosynthesis. The selection and fixation of the implant plate are vital for stabilizing the fracture. This selection is highly dependent on the plate's stability, which is influenced by factors like the stresses generated in the plate due to the load on the bone, as well as the plate's length, thickness, and number of screw holes. Minimizing these stresses is essential to reduce the risk of implant failure, ensuring optimal stress distribution and promoting faster, more effective bone healing. In the present work, the finite element and statistical approach was used to optimize the geometrical parameters of the implant plate made of SS 316L steel and Ti6Al4V alloy. A 3D finite element model was developed for analyzing the stresses and deformation, and implant plates were manufactured to validate the results with the help of an experiment conducted on the universal testing machine. A strong correlation was observed between the experimental and predicted results, with an average error of 8.6% and 8.55% for SS316L and Ti6Al4V alloy, respectively. Further, using the signal-to-noise ratio for the minimum stress condition was applied to identify the optimum parameters of the plate. Finally, regression models were developed to predict the stresses generated in SS316L and Ti6Al4V alloy plates with different input conditions. The statistical model helps us to develop the relation between different geometrical parameters of the Tibia implant plate. As determined by the present work, the parameter most influencing is implant plate length. This outcome will be used to select the implant for a specific patient, resulting in a reduction in implant failure post-surgery.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-09-27
DOI
https://doi.org/10.1088/2057-1976/ad8095
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
1.15
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Finite element analysis and optimization studies on tibia implant of SS 316L steel and Ti6Al4V alloy

Umesh Potdar, Ishan R. Sathone
2 citations
Biomedical Physics & Engineering Express
Orthopaedic implants and arthroplasty
1.15
article

Finite element analysis and optimization studies on tibia implant of SS 316L steel and Ti6Al4V alloy

Umesh Potdar, Ishan R. Sathone
article en
2 citations

Abstract

Tibial fractures account for approximately 15% of all fractures, typically resulting from high-energy trauma. A critical surgical approach to treat these fractures involves the fixation of the tibia using a plate with minimally invasive osteosynthesis. The selection and fixation of the implant plate are vital for stabilizing the fracture. This selection is highly dependent on the plate's stability, which is influenced by factors like the stresses generated in the plate due to the load on the bone, as well as the plate's length, thickness, and number of screw holes. Minimizing these stresses is essential to reduce the risk of implant failure, ensuring optimal stress distribution and promoting faster, more effective bone healing. In the present work, the finite element and statistical approach was used to optimize the geometrical parameters of the implant plate made of SS 316L steel and Ti6Al4V alloy. A 3D finite element model was developed for analyzing the stresses and deformation, and implant plates were manufactured to validate the results with the help of an experiment conducted on the universal testing machine. A strong correlation was observed between the experimental and predicted results, with an average error of 8.6% and 8.55% for SS316L and Ti6Al4V alloy, respectively. Further, using the signal-to-noise ratio for the minimum stress condition was applied to identify the optimum parameters of the plate. Finally, regression models were developed to predict the stresses generated in SS316L and Ti6Al4V alloy plates with different input conditions. The statistical model helps us to develop the relation between different geometrical parameters of the Tibia implant plate. As determined by the present work, the parameter most influencing is implant plate length. This outcome will be used to select the implant for a specific patient, resulting in a reduction in implant failure post-surgery.

Biomedical Physics & Engineering ExpressVol. 10(6)
Dr D Y Patil Dental College & Hospital (IN)
Affordable and clean energy
Openalex Percentile: Top 20%
Orthopaedic implants and arthroplasty
1.15
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.