Design and manufacture of an ultra-compact, 1.5 T class, controlled-contact resistance, REBCO, brain imaging MRI magnet

Brain imaging MRI comprises a significant proportion of MRI scans, but the requirement for including the shoulders in the magnet bore means there is not a significant size reduction in the magnet compared to whole-body magnets. Here we present a new design approach for brain imaging MRI magnets targeting ±20 kHz <i>B</i> <sub>0</sub> variation over the imaging volume rather than the more usual ±200 Hz making use of novel high-bandwidth MRI pulse sequences and distortion correction. Using this design approach, we designed and manufactured a 1.5 T class ReBCO cryogen-free magnet. The magnet is dome-like in form, completely excludes the shoulders and is <400 mm long. The magnet was wound using no-insulation style coils with a conductive epoxy encapsulant where the contact resistance of the coils was controlled so the emergency shut-down time of the magnet was less than 30 s. Despite acceptable coil testing results ahead of manufacture, during testing of the magnet, several of the epoxy coils showed signs of damage limiting stable performance to <55 A compared to the designed 160 A. These coils were replaced with insulated paraffin encapsulated coils. Subsequently the magnet was re-ramped and was stable at 81 A, generating 0.71 T as several other coils had sustained damage not visible in the first magnet iteration. The magnet has been passive shimmed to ±20 kHz <i>B</i> <sub>0</sub> variation over the imaging volume and integrated into an MRI scanner. The stability of the magnet has been evaluated and found to be acceptable for MRI.

Authors

Institutions

Publication Details

Journal
Superconductor Science and Technology
Published
2024-09-27
DOI
https://doi.org/10.1088/1361-6668/ad80d5
Citations
7
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
2.83

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and manufacture of an ultra-compact, 1.5 T class, controlled-contact resistance, REBCO, brain imaging MRI magnet

Michael Garwood, Taylor Froelich, J Bailey, Jamal R. Olatunji et al.
7 citations
Superconductor Science and Technology
Advanced MRI Techniques and Applications
2.83
article

Design and manufacture of an ultra-compact, 1.5 T class, controlled-contact resistance, REBCO, brain imaging MRI magnet

Michael Garwood, Taylor Froelich, J Bailey, Jamal R. Olatunji, H.W. Weijers, Mathieu Szmigiel, Mark Hunter, Ben Parkinson, Konstantinos Bouloukakis
article en
7 citations

Abstract

Brain imaging MRI comprises a significant proportion of MRI scans, but the requirement for including the shoulders in the magnet bore means there is not a significant size reduction in the magnet compared to whole-body magnets. Here we present a new design approach for brain imaging MRI magnets targeting ±20 kHz <i>B</i> <sub>0</sub> variation over the imaging volume rather than the more usual ±200 Hz making use of novel high-bandwidth MRI pulse sequences and distortion correction. Using this design approach, we designed and manufactured a 1.5 T class ReBCO cryogen-free magnet. The magnet is dome-like in form, completely excludes the shoulders and is <400 mm long. The magnet was wound using no-insulation style coils with a conductive epoxy encapsulant where the contact resistance of the coils was controlled so the emergency shut-down time of the magnet was less than 30 s. Despite acceptable coil testing results ahead of manufacture, during testing of the magnet, several of the epoxy coils showed signs of damage limiting stable performance to <55 A compared to the designed 160 A. These coils were replaced with insulated paraffin encapsulated coils. Subsequently the magnet was re-ramped and was stable at 81 A, generating 0.71 T as several other coils had sustained damage not visible in the first magnet iteration. The magnet has been passive shimmed to ±20 kHz <i>B</i> <sub>0</sub> variation over the imaging volume and integrated into an MRI scanner. The stability of the magnet has been evaluated and found to be acceptable for MRI.

Superconductor Science and TechnologyVol. 37(11)
University of Minnesota (US), Victoria University of Wellington (NZ), Air Liquide (France) (FR), Transpower (New Zealand) (NZ), Resonance Research (United States) (US), Open Polytechnic (NZ)
National Institutes of Health, Goddard Space Flight Center, National Institute of Biomedical Imaging and Bioengineering
Openalex Percentile: Top 9%
Advanced MRI Techniques and Applications
2.83
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.