Development of an Automatic Liquid Dosing System in Microliter Scale

The goal of this study is to design a novel automatic liquid dosing system for liquid sampling at the microliter level. For this purpose, a mechatronics system is designed to position a syringe at the desired position in the workspace and then drive its piston to inject the liquid to be sampled. Then, an application-specific algorithm is developed to be able to prepare samples in 3x3, 4x4, and 5x5 sample container arrays with a desired volume. The performance tests are conducted for preparing samples with up to three different liquids. The repetitive experiments are performed for 50 and 100 µL sampling volumes. The results indicated that it is possible to dose a single liquid with the highest average deviation of 3.9%. Moreover, it is found that it is possible to prepare a sample with a mixture of three liquids by the highest average deviation from the reference value of around 3.4% when the targeting sampling volume is 250 µL for each liquid.

Authors

Institutions

Publication Details

Journal
Deu Muhendislik Fakultesi Fen ve Muhendislik
Published
2024-09-27
DOI
https://doi.org/10.21205/deufmd.2024267819
Primary Topic
Innovative Microfluidic and Catalytic Techniques Innovation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of an Automatic Liquid Dosing System in Microliter Scale

Serkan Doğanay, Haydar Barış Öcal, Ayberk Kıllı, Ömer Yüce et al.
Deu Muhendislik Fakultesi Fen ve Muhendislik
Innovative Microfluidic and Catalytic Techniques Innovation
article

Development of an Automatic Liquid Dosing System in Microliter Scale

Serkan Doğanay, Haydar Barış Öcal, Ayberk Kıllı, Ömer Yüce, Kadri Emre Orgun
article en

Abstract

The goal of this study is to design a novel automatic liquid dosing system for liquid sampling at the microliter level. For this purpose, a mechatronics system is designed to position a syringe at the desired position in the workspace and then drive its piston to inject the liquid to be sampled. Then, an application-specific algorithm is developed to be able to prepare samples in 3x3, 4x4, and 5x5 sample container arrays with a desired volume. The performance tests are conducted for preparing samples with up to three different liquids. The repetitive experiments are performed for 50 and 100 µL sampling volumes. The results indicated that it is possible to dose a single liquid with the highest average deviation of 3.9%. Moreover, it is found that it is possible to prepare a sample with a mixture of three liquids by the highest average deviation from the reference value of around 3.4% when the targeting sampling volume is 250 µL for each liquid.

Deu Muhendislik Fakultesi Fen ve MuhendislikVol. 26(78)
Izmir Kâtip Çelebi University (TR)
Türkiye Bilimsel ve Teknolojik Araştırma Kurumu
Openalex Percentile: Top 85%
Innovative Microfluidic and Catalytic Techniques Innovation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.