Comparison of the PlusoptiX A16 and vision screener V100
The Vision Screener V100, while generally aligning well with PlusoptiX A16, tends to underestimate myopic spherical equivalent, overestimate hyperopic cases, and underestimate J0 astigmatism.
Authors
- Jorge Jorge (ORCID: https://orcid.org/0000-0002-4369-3332)
- Paulo Fernandes (ORCID: https://orcid.org/0000-0002-3921-6357)
Institutions
- University of Minho (PT)
Publication Details
- Journal
- Frontiers in Ophthalmology
- Published
- 2024-09-26
- DOI
- https://doi.org/10.3389/fopht.2024.1414417
- Citations
- 2
- Primary Topic
- Ophthalmology and Visual Impairment Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.77
Funders
- Fundação para a Ciência e a Tecnologia