Novel multi‐spectral short‐wave infrared imaging for assessment of human burn wound depth

Burn depth determination is critical for patient care but is currently lacking accuracy. Recent animal studies showed that Short Wave Infrared (SWIR) imaging can distinguish between superficial and deep burns. This is a first human study correlating reflectance of multiple SWIR bands using a SWIR assessment tool (SWAT) with burn depth classifications by surgeons and histology. Burns and adjacent normal skin in 11 patients with thermal injuries were imaged with visual and narrow bands centred at 1200, 1650, 1940 and 2250 nm and biopsies were taken from select areas. Reflectance intensities for each band in 273 regions of interest (ROI) were divided by the normal skin reflectance and combined into three Reflectance Indices (RIs). In addition, burns in ROIs and biopsies were classified by five surgeons and three pathologists, respectively, as superficial partial, deep partial, or full thickness. Results show that for burn depth increase classified by the surgeons, reflectance increased at 1200 and 2250, decreased at 1940, and didn't change at 1650 nm. In contrast, all three RIs increase with burn depth and predict the deep and full depths ROIs representing operable regions (Area Under Curve >0.6507, p < 0.0001). Pathologists' classification matched surgeons' classification of burn category only in eight of 21 biopsies (38.1%), but reflectance at all bands and one RI for all deep partial and full thickness biopsies were larger than in non-biopsy normal and superficial partial thickness ROIs (p < 0.0118). In conclusion, multi-spectral imaging with a new SWAT is a promising approach for evaluation of burn wound depth.

Authors

Institutions

Publication Details

Journal
Wound Repair and Regeneration
Published
2024-09-26
DOI
https://doi.org/10.1111/wrr.13221
Citations
5
Primary Topic
Wound Healing and Treatments
Type
article
Field-Weighted Citation Impact
1.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel multi‐spectral short‐wave infrared imaging for assessment of human burn wound depth

Brett D. Arnoldo, Bingchun Wan, Johanna Nunez, Sneha Korlakunta et al.
5 citations
Wound Repair and Regeneration
Wound Healing and Treatments
1.25
article

Novel multi‐spectral short‐wave infrared imaging for assessment of human burn wound depth

Brett D. Arnoldo, Bingchun Wan, Johanna Nunez, Sneha Korlakunta, Deborah Carlson, Jeremy Goverman, Rodney K. Chan, Chiaka Akarichi, Bret M. Evers, Samuel P. Mandell, Omer Berenfeld, Benjamin Levi, Audra T. Clark, Caroline Park, Sergey V. Mironov, Ryan M. Huebinger, Alaa Hazime
article en
5 citations

Abstract

Burn depth determination is critical for patient care but is currently lacking accuracy. Recent animal studies showed that Short Wave Infrared (SWIR) imaging can distinguish between superficial and deep burns. This is a first human study correlating reflectance of multiple SWIR bands using a SWIR assessment tool (SWAT) with burn depth classifications by surgeons and histology. Burns and adjacent normal skin in 11 patients with thermal injuries were imaged with visual and narrow bands centred at 1200, 1650, 1940 and 2250 nm and biopsies were taken from select areas. Reflectance intensities for each band in 273 regions of interest (ROI) were divided by the normal skin reflectance and combined into three Reflectance Indices (RIs). In addition, burns in ROIs and biopsies were classified by five surgeons and three pathologists, respectively, as superficial partial, deep partial, or full thickness. Results show that for burn depth increase classified by the surgeons, reflectance increased at 1200 and 2250, decreased at 1940, and didn't change at 1650 nm. In contrast, all three RIs increase with burn depth and predict the deep and full depths ROIs representing operable regions (Area Under Curve >0.6507, p < 0.0001). Pathologists' classification matched surgeons' classification of burn category only in eight of 21 biopsies (38.1%), but reflectance at all bands and one RI for all deep partial and full thickness biopsies were larger than in non-biopsy normal and superficial partial thickness ROIs (p < 0.0118). In conclusion, multi-spectral imaging with a new SWAT is a promising approach for evaluation of burn wound depth.

Wound Repair and RegenerationVol. 32(6)
Brooke Army Medical Center (US), University of Michigan (US), Massachusetts General Hospital (US), Organogenesis (United States) (US), The University of Texas Southwestern Medical Center (US)
U.S. Department of Defense, University of Texas Southwestern Medical Center, National Institutes of Health
Openalex Percentile: Top 18%
Wound Healing and Treatments
1.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.