A self-established method for the quantitative determination of plasma free hemoglobin utilizing the hemolysis index of Siemens ADVIA 2400 chemistry system

Objectives: The significance of the quantitative hemolysis index (HI) obtained from the Siemens ADVIA 2400 Chemistry System remains undisclosed. This study aims to establish a method utilizing HI of Siemens ADVIA 2400 Chemistry System for the quantitative determination of plasma free hemoglobin (fHb). Materials and Methods: The performance of the self-established method was assessed following Clinical and Laboratory Standards Institute protocols, which included detection capability, linearity, imprecision, and trueness, using materials derived from residual whole blood samples. Statistical Analysis: The analysis of variance was performed using Microsoft Excel, and least-squares linear regression analysis was conducted using MedCalc (Version 20.216). A significance level of P < 0.05 was considered. Results: The repeatability and within-laboratory precision both exhibited a coefficient of variation below 1%. The limit of blank was determined to be 0.005 g/L, while the lower limit of detection and lower limit of quantification were found to be 0.050 g/L and 0.085 g/L, respectively. Our self-established method for determining plasma fHb demonstrated linearity across the entire assay range of 0.085~25.2 g/L with an excellent correlation coefficient (R 2 = 0.9998). In the method comparison study between Siemens ADVIA 2400 and Sysmex XN-1000, our results indicated a slight bias, with only 5% (2/40) of the plots falling outside the 95% confidence interval. This suggests a high level of agreement observed between these two devices. Conclusions: The self-established method was evaluated as a reliable assay with robust performance for quantifying minute quantities of plasma fHb in the ADVIA 2400 chemistry system. Calibration for the HI was deemed necessary when determining plasma fHb.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Physicians
Published
2024-09-26
DOI
https://doi.org/10.25259/jlp_130_2024
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A self-established method for the quantitative determination of plasma free hemoglobin utilizing the hemolysis index of Siemens ADVIA 2400 chemistry system

Haifeng Xie, Jinlong Wei, Xianzhong Luo
Journal of Laboratory Physicians
Clinical Laboratory Practices and Quality Control
article

A self-established method for the quantitative determination of plasma free hemoglobin utilizing the hemolysis index of Siemens ADVIA 2400 chemistry system

Haifeng Xie, Jinlong Wei, Xianzhong Luo
article en

Abstract

Objectives: The significance of the quantitative hemolysis index (HI) obtained from the Siemens ADVIA 2400 Chemistry System remains undisclosed. This study aims to establish a method utilizing HI of Siemens ADVIA 2400 Chemistry System for the quantitative determination of plasma free hemoglobin (fHb). Materials and Methods: The performance of the self-established method was assessed following Clinical and Laboratory Standards Institute protocols, which included detection capability, linearity, imprecision, and trueness, using materials derived from residual whole blood samples. Statistical Analysis: The analysis of variance was performed using Microsoft Excel, and least-squares linear regression analysis was conducted using MedCalc (Version 20.216). A significance level of P < 0.05 was considered. Results: The repeatability and within-laboratory precision both exhibited a coefficient of variation below 1%. The limit of blank was determined to be 0.005 g/L, while the lower limit of detection and lower limit of quantification were found to be 0.050 g/L and 0.085 g/L, respectively. Our self-established method for determining plasma fHb demonstrated linearity across the entire assay range of 0.085~25.2 g/L with an excellent correlation coefficient (R 2 = 0.9998). In the method comparison study between Siemens ADVIA 2400 and Sysmex XN-1000, our results indicated a slight bias, with only 5% (2/40) of the plots falling outside the 95% confidence interval. This suggests a high level of agreement observed between these two devices. Conclusions: The self-established method was evaluated as a reliable assay with robust performance for quantifying minute quantities of plasma fHb in the ADVIA 2400 chemistry system. Calibration for the HI was deemed necessary when determining plasma fHb.

Journal of Laboratory PhysiciansVol. 16
Zhejiang Chinese Medical University (CN), Dalian Municipal Central Hospital (CN)
Siemens Healthineers
Openalex Percentile: Top 77%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.