X-ray spectral performance of the Sony IMX290 CMOS sensor near Fano limit after a per-pixel gain calibration

The advent of back-illuminated complementary metal–oxide–semiconductor (CMOS) sensors and their well-known advantages over charge-coupled devices make them an attractive technology for future X-ray missions. However, numerous challenges remain, including improving their depletion depth and identifying effective methods to calculate per-pixel gain conversion. We have tested a commercial Sony IMX290LLR CMOS sensor under X-ray light using an Fe55 radioactive source and collected X-ray photons for ∼15 consecutive days under stable conditions at regulated temperatures of 21°C and 26°C. At each temperature, the data set contained enough X-ray photons to produce one spectrum per pixel consisting only of single-pixel events. We determined the gain dispersion of its 2.1 million pixels using the peak fitting and the energy calibration via correlation (ECC) methods. We measured a gain dispersion of 0.4% at both temperatures and demonstrated the advantage of the ECC method in the case of spectra with low statistics. The energy resolution at 5.9 keV after the per-pixel gain correction is improved by ≳10 eV for single-pixel and all event spectra, with single-pixel event energy resolution reaching 123.6±0.2 eV, close to the Fano limit of silicon sensors at room temperature. Finally, our long data acquisition demonstrated the excellent stability of the detector over more than 30 days under a flux of 104 photons per second.

Authors

Institutions

Publication Details

Journal
Journal of Astronomical Telescopes Instruments and Systems
Published
2024-09-28
DOI
https://doi.org/10.1117/1.jatis.10.3.038001
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.18
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

X-ray spectral performance of the Sony IMX290 CMOS sensor near Fano limit after a per-pixel gain calibration

O. Limousin, Benjamin Schneider, Beverly LaMarr, Eric D. Miller et al.
1 citations
Journal of Astronomical Telescopes Instruments and Systems
Advanced X-ray and CT Imaging
0.18
article

X-ray spectral performance of the Sony IMX290 CMOS sensor near Fano limit after a per-pixel gain calibration

O. Limousin, Benjamin Schneider, Beverly LaMarr, Eric D. Miller, G. Prigozhin, Nathan P. Lourie, Richard F. Foster, Andrew Malonis, Jill Juneau, Catherine E. Grant, Marshall W. Bautz, Sarah Heine, Hope Fu
article en
1 citations

Abstract

The advent of back-illuminated complementary metal–oxide–semiconductor (CMOS) sensors and their well-known advantages over charge-coupled devices make them an attractive technology for future X-ray missions. However, numerous challenges remain, including improving their depletion depth and identifying effective methods to calculate per-pixel gain conversion. We have tested a commercial Sony IMX290LLR CMOS sensor under X-ray light using an Fe55 radioactive source and collected X-ray photons for ∼15 consecutive days under stable conditions at regulated temperatures of 21°C and 26°C. At each temperature, the data set contained enough X-ray photons to produce one spectrum per pixel consisting only of single-pixel events. We determined the gain dispersion of its 2.1 million pixels using the peak fitting and the energy calibration via correlation (ECC) methods. We measured a gain dispersion of 0.4% at both temperatures and demonstrated the advantage of the ECC method in the case of spectra with low statistics. The energy resolution at 5.9 keV after the per-pixel gain correction is improved by ≳10 eV for single-pixel and all event spectra, with single-pixel event energy resolution reaching 123.6±0.2 eV, close to the Fano limit of silicon sensors at room temperature. Finally, our long data acquisition demonstrated the excellent stability of the detector over more than 30 days under a flux of 104 photons per second.

Journal of Astronomical Telescopes Instruments and SystemsVol. 10(03)
Centre National de la Recherche Scientifique (FR), Université Paris Cité (FR), Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), Université Paris-Saclay (FR), Astrophysique, Instrumentation et Modélisation (FR), CEA Paris-Saclay (FR), Massachusetts Institute of Technology (US)
Affordable and clean energy
Openalex Percentile: Top 52%
Advanced X-ray and CT Imaging
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.