Improvement of the energy resolution and non-proportional light response of CsI:Tl to gamma radiation using dynamic integration times
Abstract The analysis of the light response of thallium-doped cesium iodide to gamma radiation was performed on single scintillation waveforms recorded using a digital oscilloscope. As an alternative to classical pulse processing procedures, the use of dynamic integration times instead of fixed shaping/integration times was demonstrated. The dynamic integration time, changeable with the amplitudes of the CsI:Tl scintillation pulses, allowed to optimise the light collection with simultaneous minimization of a noise contribution. It was shown that for low-energy γ -rays (22 keV–60 keV), the scintillation pulses should be integrated using shorter integration times (approximately 7 μ s–14 μs), whereas for higher energies (662 keV–835 keV), the best energy resolution was achieved with integration times of about 28 μ s–30 μ s. The approach also allowed for the registration of more proportional CsI:Tl light responses to deposited X and gamma radiation in the energy range of 22 keV–835 keV.
Authors
- A. Gektin (ORCID: https://orcid.org/0000-0002-1733-0343)
- A. Syntfeld-Każuch (ORCID: https://orcid.org/0000-0002-8071-7380)
- Z. Mianowska (ORCID: https://orcid.org/0000-0002-5803-1825)
- S. Mianowski (ORCID: https://orcid.org/0000-0003-2514-6156)
- R. Alberts
Institutions
- National Academy of Sciences of Ukraine (UA)
- Diamond Light Source (GB)
- National Centre for Nuclear Research (PL)
- Institute for Scintillation Materials (UA)
Publication Details
- Journal
- Journal of Instrumentation
- Published
- 2024-09-01
- DOI
- https://doi.org/10.1088/1748-0221/19/09/p09029
- Citations
- 1
- Primary Topic
- Radiation Detection and Scintillator Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.61