3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN
Aiming at the problems of various nonlinear characteristics caused by fiber twist in the manufacturing and packaging process of interventional catheter shape sensor, this paper proposes a fiber grating shape sensor configuration and its packaging method. The HPO-BPNN algorithm is employed to predict discrete curvature and the direction angles of curvature vectors, thus avoiding complex sensor parameter calibration steps. Additionally, the Frenet-Serret equation is integrated to achieve three-dimensional posture reconstruction of the catheter. Through the multi-curvature calibration experiment of the designed sensor, the average error at the end of the 2D curves is 21.15mm (3.91%), and for the 3D curves, it is 26.43mm (4.89%), thereby verifying the feasibility of the data-driven shape reconstruction algorithm.
Authors
- Zhenzhen Song (ORCID: https://orcid.org/0009-0005-3014-2223)
- Tianliang Li (ORCID: https://orcid.org/0000-0002-7323-169X)
- Bin Wang (ORCID: https://orcid.org/0000-0001-5265-1030)
- Yongwen Zhu
Institutions
- Wuhan University of Technology (CN)
- Wuhan University (CN)
- Zhongnan Hospital of Wuhan University (CN)
Publication Details
- Published
- 2024-08-08
- DOI
- https://doi.org/10.1109/cis-ram61939.2024.10673189
- Citations
- 1
- Primary Topic
- Medical Image Segmentation Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.25
Funders
- National Natural Science Foundation of China