3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN

Aiming at the problems of various nonlinear characteristics caused by fiber twist in the manufacturing and packaging process of interventional catheter shape sensor, this paper proposes a fiber grating shape sensor configuration and its packaging method. The HPO-BPNN algorithm is employed to predict discrete curvature and the direction angles of curvature vectors, thus avoiding complex sensor parameter calibration steps. Additionally, the Frenet-Serret equation is integrated to achieve three-dimensional posture reconstruction of the catheter. Through the multi-curvature calibration experiment of the designed sensor, the average error at the end of the 2D curves is 21.15mm (3.91%), and for the 3D curves, it is 26.43mm (4.89%), thereby verifying the feasibility of the data-driven shape reconstruction algorithm.

Authors

Institutions

Publication Details

Published
2024-08-08
DOI
https://doi.org/10.1109/cis-ram61939.2024.10673189
Citations
1
Primary Topic
Medical Image Segmentation Techniques
Type
article
Field-Weighted Citation Impact
0.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN

Zhenzhen Song, Tianliang Li, Bin Wang, Yongwen Zhu
1 citations
Medical Image Segmentation Techniques
0.25
article

3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN

Zhenzhen Song, Tianliang Li, Bin Wang, Yongwen Zhu
article en
1 citations

Abstract

Aiming at the problems of various nonlinear characteristics caused by fiber twist in the manufacturing and packaging process of interventional catheter shape sensor, this paper proposes a fiber grating shape sensor configuration and its packaging method. The HPO-BPNN algorithm is employed to predict discrete curvature and the direction angles of curvature vectors, thus avoiding complex sensor parameter calibration steps. Additionally, the Frenet-Serret equation is integrated to achieve three-dimensional posture reconstruction of the catheter. Through the multi-curvature calibration experiment of the designed sensor, the average error at the end of the 2D curves is 21.15mm (3.91%), and for the 3D curves, it is 26.43mm (4.89%), thereby verifying the feasibility of the data-driven shape reconstruction algorithm.

Wuhan University of Technology (CN), Wuhan University (CN), Zhongnan Hospital of Wuhan University (CN)
National Natural Science Foundation of China
Sustainable cities and communities
Openalex Percentile: Top 47%
Medical Image Segmentation Techniques
0.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN — Zhenzhen Song, Tianliang Li, et al. · (2024) | TGRS Research Map | TGRS