PWM-Based Control Circuit for Fast Heating of Flexible Microheaters

This paper presents the design and development of heater control for a flexible embedded microheater (7000 /Lm x 9000 /Lm) for a lab-on-a-chip system. Our work focuses on advancing the PWM-based control system with the goal of reducing the system time constant. We also present the fabrication of an embedded flexible microheater in a PDMS substrate. We validate the efficacy of the flexible microheater and its control system by using it for glucose detection. A steady state temperature in the range of 40° C to 120° C is achieved in approximately 40 s, with the heater's power consumption ranging from 0.4 W to 2 W. The novelty of the work lies in that we have made a small, flexible, low-cost chip without the use of sophisticated equipment or conventional fabrication techniques.

Authors

Institutions

Publication Details

Published
2024-08-11
DOI
https://doi.org/10.1109/mwscas60917.2024.10658773
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

PWM-Based Control Circuit for Fast Heating of Flexible Microheaters

Anshu Sarje, Mst Shamim Ara Shawkat, Anjali Singh
Advanced Sensor and Energy Harvesting Materials
article

PWM-Based Control Circuit for Fast Heating of Flexible Microheaters

Anshu Sarje, Mst Shamim Ara Shawkat, Anjali Singh
article en

Abstract

This paper presents the design and development of heater control for a flexible embedded microheater (7000 /Lm x 9000 /Lm) for a lab-on-a-chip system. Our work focuses on advancing the PWM-based control system with the goal of reducing the system time constant. We also present the fabrication of an embedded flexible microheater in a PDMS substrate. We validate the efficacy of the flexible microheater and its control system by using it for glucose detection. A steady state temperature in the range of 40° C to 120° C is achieved in approximately 40 s, with the heater's power consumption ranging from 0.4 W to 2 W. The novelty of the work lies in that we have made a small, flexible, low-cost chip without the use of sophisticated equipment or conventional fabrication techniques.

Florida International University (US), Indian Institute of Technology Hyderabad (IN)
Affordable and clean energy
Openalex Percentile: Top 86%
Advanced Sensor and Energy Harvesting Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.