Evaluation of the Oncomine Comprehensive Assay Plus NGS Panel and the OncoScan CNV Assay for Homologous Recombination Deficiency Detection
The OCA Plus assay and the OncoScan CNV assay show a high but not complete concordance to reference standard homologous recombination deficiency (HRD) detection. The main reason for QC failure or non-concordance in our study was a low tumor fraction estimated in the assay, despite the selection of material by a pathologist with an inclusion criterion of > 30% tumor. QC steps should include careful tumor content evaluation, and results on samples with < 30% tumor should not be reported.