Evaluation of an automated synthesis method for [18F]Fluoromisonidazole using a Scintomics GRP® module
Abstract There is limited information on utilizing commercially supplied [ 18 F]Fluoride from an off-site cyclotron for the synthesis of radiopharmaceuticals. This study explored the production of the PET hypoxia marker [ 18 F]FMISO, using the Scintomics® GRP module and distantly produced [ 18 F]Fluoride, which lacks published data. A radiochemical yield of 27.4 ± 3.6% ( n = 5) and high radiochemical purity were achieved in synthesis time of 48 min. The [ 18 F]FMISO met all Ph. Eur. standards, demonstrating the consistency and reliability of the module in generating a clinical-grade radiopharmaceutical using off-site produced [ 18 F]Fluoride, but available patient doses were severely limited.
Authors
- Sietske Rubow (ORCID: https://orcid.org/0000-0001-8194-9848)
- Jannie le Roux (ORCID: https://orcid.org/0000-0001-5075-3236)
- J. Eric Wood Oliver (ORCID: https://orcid.org/0000-0002-9906-421X)
Institutions
- Stellenbosch University (ZA)
Publication Details
- Journal
- Journal of Radioanalytical and Nuclear Chemistry
- Published
- 2024-09-21
- DOI
- https://doi.org/10.1007/s10967-024-09734-w
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Universiteit Stellenbosch