Development of a superheated water chromatography method for the sustainable determination of preservatives in foods and cosmetics

• Porous graphitic carbon columns allowed separations at extreme temperatures (up to 250 °C). • The retention decrease at high temperature enabled analysis of parabens at reduced analysis time. • The developed method was validated in terms of linearity, accuracy, precision, LOD and LOQ. • Eco-Scales, AGREE and GAPI tools were employed for greenness evaluation. • The proposed method was greener than current official method for the analysis of parabens. The present research focused on the use of high temperatures to achieve fast separations in Liquid Chromatography (LC) and minimize the consumption of organic solvents. Particularly, superheated water was employed, by exploiting the decrease in dielectric constant of water when increasing the temperature. To realize this kind of application, the availability of LC columns resistant at high temperature is mandatory. Recently, porous graphitic carbon (PGC) columns have been proposed as stationary phases resistant up to 250 °C. Hardware modifications were also necessary to guarantee the efficient heating of the mobile phase prior of entering the column, as well as its rapid cooling before detection by photodiode array (PDA). Specifically, an LC system was interfaced to a GC oven, which hosted the LC column; to achieve a fast and efficient heating of the eluent coming from the autosampler prior to enter into the column, a pre-heating tube was interposed between the autosampler outlet and the column inlet. Instead, a cooling loop was connected between the column outlet and the PDA cell. A mixture of 1-nitroalkanes was used to evaluate the performance of the system in terms of dead volumes, band broadening and peak shape, as well as chromatographic efficiency and resolution. The use of higher temperatures allowed to perform analyses at higher flow rates in reduced analysis time and was also in terms of peak bandwidth resulting to increased efficiency for the most retained compound. Moreover, van't Hoff plots were obtained at different temperature, confirming that the most retained analyte greatly benefited of the temperature increase. The system was applied to the analysis of parabens in cosmetic and food products. The method combined eco-sustainable sample preparation with high-temperature LC separation on a PGC column and was fully validated in terms of linearity, precision, accuracy and detection and quantification limits. The developed method was critically evaluated by comparing the output of different metric tools implemented in the last decade for the quantitative assessment of the greenness.

Authors

Institutions

Publication Details

Journal
Green Analytical Chemistry
Published
2024-09-21
DOI
https://doi.org/10.1016/j.greeac.2024.100155
Citations
1
Primary Topic
Chemistry and Chemical Engineering
Type
article
Field-Weighted Citation Impact
0.25
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a superheated water chromatography method for the sustainable determination of preservatives in foods and cosmetics

Francesca Rigano, Luigi Mondello, Patrik Appelblad, Roberta La Tella et al.
1 citations
Green Analytical Chemistry
Chemistry and Chemical Engineering
0.25
article

Development of a superheated water chromatography method for the sustainable determination of preservatives in foods and cosmetics

Francesca Rigano, Luigi Mondello, Patrik Appelblad, Roberta La Tella, Paola Dugo, Danilo Donnarumma, Alice Mondello, Paola Donato
article en
1 citations

Abstract

• Porous graphitic carbon columns allowed separations at extreme temperatures (up to 250 °C). • The retention decrease at high temperature enabled analysis of parabens at reduced analysis time. • The developed method was validated in terms of linearity, accuracy, precision, LOD and LOQ. • Eco-Scales, AGREE and GAPI tools were employed for greenness evaluation. • The proposed method was greener than current official method for the analysis of parabens. The present research focused on the use of high temperatures to achieve fast separations in Liquid Chromatography (LC) and minimize the consumption of organic solvents. Particularly, superheated water was employed, by exploiting the decrease in dielectric constant of water when increasing the temperature. To realize this kind of application, the availability of LC columns resistant at high temperature is mandatory. Recently, porous graphitic carbon (PGC) columns have been proposed as stationary phases resistant up to 250 °C. Hardware modifications were also necessary to guarantee the efficient heating of the mobile phase prior of entering the column, as well as its rapid cooling before detection by photodiode array (PDA). Specifically, an LC system was interfaced to a GC oven, which hosted the LC column; to achieve a fast and efficient heating of the eluent coming from the autosampler prior to enter into the column, a pre-heating tube was interposed between the autosampler outlet and the column inlet. Instead, a cooling loop was connected between the column outlet and the PDA cell. A mixture of 1-nitroalkanes was used to evaluate the performance of the system in terms of dead volumes, band broadening and peak shape, as well as chromatographic efficiency and resolution. The use of higher temperatures allowed to perform analyses at higher flow rates in reduced analysis time and was also in terms of peak bandwidth resulting to increased efficiency for the most retained compound. Moreover, van't Hoff plots were obtained at different temperature, confirming that the most retained analyte greatly benefited of the temperature increase. The system was applied to the analysis of parabens in cosmetic and food products. The method combined eco-sustainable sample preparation with high-temperature LC separation on a PGC column and was fully validated in terms of linearity, precision, accuracy and detection and quantification limits. The developed method was critically evaluated by comparing the output of different metric tools implemented in the last decade for the quantitative assessment of the greenness.

Green Analytical ChemistryVol. 11
University of Messina (IT), MSD (Norway) (NO)
Zero hunger
Openalex Percentile: Top 43%
Chemistry and Chemical Engineering
0.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.