Unsupervised Retrospective Detection of Pressure Induced Failures in Continuous Glucose Monitoring Sensors for T1D Management

Continuous Glucose Monitoring sensors (CGMs) have revolutionized type 1 diabetes (T1D) management. In particular, in several cases, the retrospective analysis of CGM recordings allows clinicians to review and adjust patients' therapy. However, in this set-up, the artifacts that are often present in CGM data could lead to incorrect therapeutic actions. To mitigate this risk, we investigate how to detect one of the most common of these artifacts, the so-called pressure induced sensor attenuations, by means of anomaly detection algorithms. Specifically, these methods belong to the class of unsupervised techniques, which is particularly appealing since it does not require a labeled dataset, hardly available in practice. After having designed five features to highlight the anomalous state of the sensor, 8 different methods (e.g. Isolation Forest and Histogram-based Outlier Score) are assessed both in silico using the UVa/Padova Type 1 Diabetes Simulator and on real data of 36 subjects monitored for about 10 days. In the in silico scenario, the best results are achieved with Isolation Forest, which recognized the 74% of the failures generating on average only 2 false alerts during the whole monitoring time. In real data, Isolation Forest is confirmed to be effective in the detection of failures, achieving a recall of 55% and generating 3 false alarms in 10 days. By allowing to detect more than 50% of the artifacts while discarding only a few portions of correct data in several days of monitoring, the proposed approach could effectively improve the quality of CGM data used by clinicians to retrospectively evaluate and adjust T1D therapy.

Authors

Institutions

Publication Details

Journal
IEEE Journal of Biomedical and Health Informatics
Published
2024-09-20
DOI
https://doi.org/10.1109/jbhi.2024.3465893
Citations
3
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
1.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Unsupervised Retrospective Detection of Pressure Induced Failures in Continuous Glucose Monitoring Sensors for T1D Management

Simone Del Favero, Eleonora Manzoni, Elena Idi, Giovanni Sparacino et al.
3 citations
IEEE Journal of Biomedical and Health Informatics
Diabetes Management and Research
1.07
article

Unsupervised Retrospective Detection of Pressure Induced Failures in Continuous Glucose Monitoring Sensors for T1D Management

Simone Del Favero, Eleonora Manzoni, Elena Idi, Giovanni Sparacino, Andrea Facchinetti
article en
3 citations

Abstract

Continuous Glucose Monitoring sensors (CGMs) have revolutionized type 1 diabetes (T1D) management. In particular, in several cases, the retrospective analysis of CGM recordings allows clinicians to review and adjust patients' therapy. However, in this set-up, the artifacts that are often present in CGM data could lead to incorrect therapeutic actions. To mitigate this risk, we investigate how to detect one of the most common of these artifacts, the so-called pressure induced sensor attenuations, by means of anomaly detection algorithms. Specifically, these methods belong to the class of unsupervised techniques, which is particularly appealing since it does not require a labeled dataset, hardly available in practice. After having designed five features to highlight the anomalous state of the sensor, 8 different methods (e.g. Isolation Forest and Histogram-based Outlier Score) are assessed both in silico using the UVa/Padova Type 1 Diabetes Simulator and on real data of 36 subjects monitored for about 10 days. In the in silico scenario, the best results are achieved with Isolation Forest, which recognized the 74% of the failures generating on average only 2 false alerts during the whole monitoring time. In real data, Isolation Forest is confirmed to be effective in the detection of failures, achieving a recall of 55% and generating 3 false alarms in 10 days. By allowing to detect more than 50% of the artifacts while discarding only a few portions of correct data in several days of monitoring, the proposed approach could effectively improve the quality of CGM data used by clinicians to retrospectively evaluate and adjust T1D therapy.

IEEE Journal of Biomedical and Health InformaticsVol. 29(2)
University of Padua (IT)
Openalex Percentile: Top 20%
Diabetes Management and Research
1.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Unsupervised Retrospective Detection of Pressure Induced Failures in Continuous Glucose Monitoring Sensors for T1D Management — Simone Del Favero, Eleonora Manzoni, et al. · IEEE Journal of Biomedical and Health Informatics (2024) | TGRS Research Map | TGRS