A case report of critical aortic stenosis diagnosed utilizing non-imaging continuous wave Doppler probe
In our specific case, dynamic acoustic shadowing of the aortic valve from the apical window obscured both imaging and Doppler signals. This acoustic shadowing was not present from the right sternal border with the NI-CWD probe, leading to an over 100% increase in aortic valve peak velocity and proper correction of AS severity. This allowed for expedited care and underscores the importance of such techniques.