Characterization of the response of radiochromic film to quasi-monoenergetic x rays through a cross-calibration with image plates

Radiochromic film (RCF) and image plates (IPs) are both commonly used detectors in diagnostics fielded at inertial confinement fusion (ICF) and high-energy-density physics (HEDP) research facilities. Due to the intense x-ray background in all ICF/HEDP experiments, accurately calibrating the optical density of RCF as a function of x-ray dose, and the photostimulated luminescence per photon of IPs as a function of x-ray energy, is necessary for interpreting experimental results. Various measurements of the sensitivity curve of different IPs to x rays have been performed [Izumi et al., Proc. SPIE 8850, 885006 (2013) and Rosenberg et al., Rev. Sci. Instrum. 90(1), 013506 (2019)]; however, calibrating RCF is a tedious process that depends on factors such as the orientation in which the RCF is scanned in the film scanner and the batch of RCF used. These issues can be mitigated by cross-calibrating RCF with IPs to enable the use of IPs for the determination of dose on the RCF without scanning the RCF. Here, the first cross-calibration of RCF with IPs to quasi-monoenergetic titanium, copper, and molybdenum K-line x rays is presented. It is found that the IP-inferred dose rates on the RCF for the Ti and Mo x rays agree well with the measured dose rates, while the IP-inferred dose rate for the Cu x rays is larger than the measured dose rate by ∼2×. Explanations for this discrepancy and plans for future work are discussed.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2024-09-01
DOI
https://doi.org/10.1063/5.0219340
Citations
1
Primary Topic
Laser-Plasma Interactions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.66

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization of the response of radiochromic film to quasi-monoenergetic x rays through a cross-calibration with image plates

Alison M. Devault, Maria Gatu Johnson, S. G. Dannhoff, B. C. Foo et al.
1 citations
Review of Scientific Instruments
Laser-Plasma Interactions and Diagnostics
0.66
article

Characterization of the response of radiochromic film to quasi-monoenergetic x rays through a cross-calibration with image plates

Alison M. Devault, Maria Gatu Johnson, S. G. Dannhoff, B. C. Foo, J. A. Pearcy, T. E. Evans, R. D. Petrasso, Y. Lawrence, J. Vargas, B. L. Reichelt, Johan A. Frenje, T. M. Johnson, B. I. Buschmann, M. Cufari, C. W. Wink, J. H. Kunimune, L. Russell, N. Vanderloo
article en
1 citations

Abstract

Radiochromic film (RCF) and image plates (IPs) are both commonly used detectors in diagnostics fielded at inertial confinement fusion (ICF) and high-energy-density physics (HEDP) research facilities. Due to the intense x-ray background in all ICF/HEDP experiments, accurately calibrating the optical density of RCF as a function of x-ray dose, and the photostimulated luminescence per photon of IPs as a function of x-ray energy, is necessary for interpreting experimental results. Various measurements of the sensitivity curve of different IPs to x rays have been performed [Izumi et al., Proc. SPIE 8850, 885006 (2013) and Rosenberg et al., Rev. Sci. Instrum. 90(1), 013506 (2019)]; however, calibrating RCF is a tedious process that depends on factors such as the orientation in which the RCF is scanned in the film scanner and the batch of RCF used. These issues can be mitigated by cross-calibrating RCF with IPs to enable the use of IPs for the determination of dose on the RCF without scanning the RCF. Here, the first cross-calibration of RCF with IPs to quasi-monoenergetic titanium, copper, and molybdenum K-line x rays is presented. It is found that the IP-inferred dose rates on the RCF for the Ti and Mo x rays agree well with the measured dose rates, while the IP-inferred dose rate for the Cu x rays is larger than the measured dose rate by ∼2×. Explanations for this discrepancy and plans for future work are discussed.

Review of Scientific InstrumentsVol. 95(9)
Plasma Technology (United States) (US), Fusion Academy (US), Massachusetts Institute of Technology (US)
National Nuclear Security Administration
Affordable and clean energy
Openalex Percentile: Top 14%
Laser-Plasma Interactions and Diagnostics
0.66
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.