Purpose: Leg blood flow has been measured using a single near-infrared spectroscopy (NIRS) device after exercise or ischemia. This study evaluated leg blood flow using three NIRS devices on three locations, using cuff ischemia and postural ischemia. The aim was to determine the optimal method and duration for these measurements. Methods: Young, healthy adults (males: n = 5 and females: n = 2) were tested on two separate days. Blood flow was measured as the half-time of recovery (T1/2) of the NIRS tissue saturation index (TSI) after leg lifts (60s, 120 s) or cuff ischemia (30s, 60s, 180 s, 300 s). NIRS devices were placed on the vastus lateralis, medial gastrocnemius, and Extensor digitorum brevis. T-tests were made to compare values between days and tests. Results: All subjects preferred the leg lifts to cuff ischemia to measure blood flow. T1/2 values for 60 seconds and 120 seconds of leg lifts were 4.6 + 0 and 4.3 + 0 for the thigh, 5.9 + 2.5 and 7.8 + 5.4 for the calf, and 6.1 + 2.6 and 8.4 + 4.1 for the foot. The values were 4.6 + 1.7, 7.0 + 3.7, and 9.6 + 6.0 for 30 seconds of ischemia. The values were 5.2 + 1.4, 7.6 + 6.3, and 9.0 + 6.2 for 60 seconds of ischemia. The values were 7.2 + 1.3, 12.0 + 5.4, and 13.8 + 5.5 for 180 seconds of ischemia. The values were 9.2 + 1.8, 15.1 + 6.2, and 16.7 + 5.4 for 300 seconds of ischemia. The mean values were not different between leg lift durations and were slower in the foot in comparison to the thigh values for all measures. There was a significant difference between thigh and foot measurements for most durations. Conclusions: Leg lifts are a subject's preferred method to measure blood flow in the leg. Normal values for the foot will be longer than for the calf and the thigh.

Authors

Institutions

Publication Details

Journal
Medicine & Science in Sports & Exercise
Published
2024-09-16
DOI
https://doi.org/10.1249/01.mss.0001052832.93707.45
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing Leg Blood Flow In Multiple Locations Using Multiple Near-infrared Spectroscopy Devices

A. C. Lunn, Kevin K. McCully, Abisola Akinbobola, Ryan Willoughby et al.
1 links
Medicine & Science in Sports & Exercise
Non-Invasive Vital Sign Monitoring
article

Assessing Leg Blood Flow In Multiple Locations Using Multiple Near-infrared Spectroscopy Devices

A. C. Lunn, Kevin K. McCully, Abisola Akinbobola, Ryan Willoughby, Zabi Dessalegn, Jalyn Boyles
article en

Abstract

Purpose: Leg blood flow has been measured using a single near-infrared spectroscopy (NIRS) device after exercise or ischemia. This study evaluated leg blood flow using three NIRS devices on three locations, using cuff ischemia and postural ischemia. The aim was to determine the optimal method and duration for these measurements. Methods: Young, healthy adults (males: n = 5 and females: n = 2) were tested on two separate days. Blood flow was measured as the half-time of recovery (T1/2) of the NIRS tissue saturation index (TSI) after leg lifts (60s, 120 s) or cuff ischemia (30s, 60s, 180 s, 300 s). NIRS devices were placed on the vastus lateralis, medial gastrocnemius, and Extensor digitorum brevis. T-tests were made to compare values between days and tests. Results: All subjects preferred the leg lifts to cuff ischemia to measure blood flow. T1/2 values for 60 seconds and 120 seconds of leg lifts were 4.6 + 0 and 4.3 + 0 for the thigh, 5.9 + 2.5 and 7.8 + 5.4 for the calf, and 6.1 + 2.6 and 8.4 + 4.1 for the foot. The values were 4.6 + 1.7, 7.0 + 3.7, and 9.6 + 6.0 for 30 seconds of ischemia. The values were 5.2 + 1.4, 7.6 + 6.3, and 9.0 + 6.2 for 60 seconds of ischemia. The values were 7.2 + 1.3, 12.0 + 5.4, and 13.8 + 5.5 for 180 seconds of ischemia. The values were 9.2 + 1.8, 15.1 + 6.2, and 16.7 + 5.4 for 300 seconds of ischemia. The mean values were not different between leg lift durations and were slower in the foot in comparison to the thigh values for all measures. There was a significant difference between thigh and foot measurements for most durations. Conclusions: Leg lifts are a subject's preferred method to measure blood flow in the leg. Normal values for the foot will be longer than for the calf and the thigh.

Medicine & Science in Sports & ExerciseVol. 56(10S)
University of Georgia (US)
Openalex Percentile: Top 86%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Assessing Leg Blood Flow In Multiple Locations Using Multiple Near-infrared Spectroscopy Devices — A. C. Lunn, Kevin K. McCully, et al. · Medicine & Science in Sports & Exercise (2024) | TGRS Research Map | TGRS