Standardized Temporary Atrial Epicardial Wire Locations Lead to Enhanced Atrial Signal Identification

<b>Background:</b> Arrhythmias, common after pediatric cardiac surgery, are associated with increased morbidity and mortality. Atrial epicardial wires (AEW) improve diagnostic accuracy but have variable pacing and sensing properties based on their location. Even so, there are no longitudinal prospective pediatric studies examining ideal placement of AEW. <b>Methods:</b> This multicenter study compared atrial amplitudes, pacing sensitivities and thresholds via AEW placed at Bachmann's Bundle (BB) and the interatrial groove near the right pulmonary veins (IGRPV) versus the surgeons' standard locations. An AtriAmp system was used to obtain an atrial ECG to calculate atrial and ventricular amplitude from atrial electrograms on the bedside monitor. Sensitivities and thresholds via a temporary pacemaker were documented. ANOVA tests with repeated measures and post-hoc pairwise comparisons were performed to compare variables within the first 24-h postoperative hours. Mixed effects linear regression models were employed to examine daily trends. <b>Results:</b> In the first 24-h following cardiac surgery, AEW at BB and IGRPV showed significantly larger atrial amplitudes than the surgeons' standard locations. In addition, there was a negative trend in atrial ECG amplitude in all AEW from postoperative days 0 to 1; however, subsequent days showed a positive mean change in atrial amplitude with largest increase seen at BB. Atrial sensing as measured by the temporary pacemaker had statistically greater atrial signal amplitude from the BB-IGRPV set in both polarities (ie, with the BB as the - or + electrode pair) as compared to the surgeons' standard locations. No difference in atrial thresholds (mA) were noted in the immediate postoperative period or over time, with a relatively low atrial threshold at all sites. <b>Conclusion:</b> Standardization of AEW at Bachmann's Bundle can yield largest atrial amplitudes by atrial ECG and highest atrial sensing parameters without compromising atrial thresholds.

Authors

Institutions

Publication Details

Journal
Journal of Intensive Care Medicine
Published
2024-09-19
DOI
https://doi.org/10.1177/08850666241279229
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Standardized Temporary Atrial Epicardial Wire Locations Lead to Enhanced Atrial Signal Identification

Jenna Torgeson, Joshua L. Hermsen, William J. Gibson, Nir Atlas et al.
Journal of Intensive Care Medicine
Cardiac Arrhythmias and Treatments
article

Standardized Temporary Atrial Epicardial Wire Locations Lead to Enhanced Atrial Signal Identification

Jenna Torgeson, Joshua L. Hermsen, William J. Gibson, Nir Atlas, Nicholas H. Von Bergen, Xiao Zhang, Adam M. Harmon
article en

Abstract

<b>Background:</b> Arrhythmias, common after pediatric cardiac surgery, are associated with increased morbidity and mortality. Atrial epicardial wires (AEW) improve diagnostic accuracy but have variable pacing and sensing properties based on their location. Even so, there are no longitudinal prospective pediatric studies examining ideal placement of AEW. <b>Methods:</b> This multicenter study compared atrial amplitudes, pacing sensitivities and thresholds via AEW placed at Bachmann's Bundle (BB) and the interatrial groove near the right pulmonary veins (IGRPV) versus the surgeons' standard locations. An AtriAmp system was used to obtain an atrial ECG to calculate atrial and ventricular amplitude from atrial electrograms on the bedside monitor. Sensitivities and thresholds via a temporary pacemaker were documented. ANOVA tests with repeated measures and post-hoc pairwise comparisons were performed to compare variables within the first 24-h postoperative hours. Mixed effects linear regression models were employed to examine daily trends. <b>Results:</b> In the first 24-h following cardiac surgery, AEW at BB and IGRPV showed significantly larger atrial amplitudes than the surgeons' standard locations. In addition, there was a negative trend in atrial ECG amplitude in all AEW from postoperative days 0 to 1; however, subsequent days showed a positive mean change in atrial amplitude with largest increase seen at BB. Atrial sensing as measured by the temporary pacemaker had statistically greater atrial signal amplitude from the BB-IGRPV set in both polarities (ie, with the BB as the - or + electrode pair) as compared to the surgeons' standard locations. No difference in atrial thresholds (mA) were noted in the immediate postoperative period or over time, with a relatively low atrial threshold at all sites. <b>Conclusion:</b> Standardization of AEW at Bachmann's Bundle can yield largest atrial amplitudes by atrial ECG and highest atrial sensing parameters without compromising atrial thresholds.

Journal of Intensive Care MedicineVol. 40(3)
Children's Mercy Hospital (US), University of Wisconsin–Madison (US), University of Missouri–Kansas City (US)
Good health and well-being
Openalex Percentile: Top 72%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.