Comparison of measurements and calculated lens power using three biometers: a Scheimpflug tomographer with partial coherence interferometry and two swept source optical coherence tomographers
The Pentacam AXL Wave, IOLMaster 700, and ANTERION can reliably provide data for IOL power calculations; however, these data are not interchangeable. In nontoric and toric eyes, 35-45% of cases recommended the same sphere IOL power, and in another 30-40%, the Pentacam and ANTERION recommended one-step higher IOL power than the IOLMaster-derived data. In targeting emmetropia, selecting the first plus IOL power is advisable when using the Pentacam and ANTERION to approximate the IOL power calculations recommended by the IOLMaster 700.