Real-time delivered dose assessment in carbon ion therapy of moving targets

<i>Objective</i>. Real-time adaptive particle therapy is being investigated as a means to maximize the treatment delivery accuracy. To react to dosimetric errors, a system for fast and reliable verification of the agreement between planned and delivered doses is essential. This study presents a clinically feasible, real-time 4D-dose reconstruction system, synchronized with the treatment delivery and motion of the patient, which can provide the necessary feedback on the quality of the delivery.<i>Approach</i>. A GPU-based analytical dose engine capable of millisecond dose calculation for carbon ion therapy has been developed and interfaced with the next generation of the dose delivery system (DDS) in use at Centro Nazionale di Adroterapia Oncologica (CNAO). The system receives the spot parameters and the motion information of the patient during the treatment and performs the reconstruction of the planned and delivered 4D-doses. After each iso-energy layer, the results are displayed on a graphical user interface by the end of the spill pause of the synchrotron, permitting verification against the reference dose. The framework has been verified experimentally at CNAO for a lung cancer case based on a virtual phantom 4DCT. The patient's motion was mimicked by a moving Ionization Chamber (IC) 2D-array.<i>Main</i><i>results</i>. For the investigated static and 4D-optimized treatment delivery cases, real-time dose reconstruction was achieved with an average pencil beam dose calculation speed up to more than one order of magnitude smaller than the spot delivery. The reconstructed doses have been benchmarked against offline log-file based dose reconstruction with the TRiP98 treatment planning system, as well as QA measurements with the IC 2D-array, where an average gamma-index passing rate (3%/3 mm) of 99.8% and 98.3%, respectively, were achieved.<i>Significance</i>. This work provides the first real-time 4D-dose reconstruction engine for carbon ion therapy. The framework integration with the CNAO DDS paves the way for a swift transition to the clinics.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-09-19
DOI
https://doi.org/10.1088/1361-6560/ad7d59
Citations
5
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
2.09
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-time delivered dose assessment in carbon ion therapy of moving targets

Lennart Volz, M. Donetti, Christian Graeff, S. Giordanengo et al.
5 citations
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
2.09
article

Real-time delivered dose assessment in carbon ion therapy of moving targets

Lennart Volz, M. Donetti, Christian Graeff, S. Giordanengo, A. Vignati, Timo Steinsberger, Félix Más Milián, Marco Durante, Cosimo Galeone, R. Sacchi, Maria Chiara Martire
article en
5 citations

Abstract

<i>Objective</i>. Real-time adaptive particle therapy is being investigated as a means to maximize the treatment delivery accuracy. To react to dosimetric errors, a system for fast and reliable verification of the agreement between planned and delivered doses is essential. This study presents a clinically feasible, real-time 4D-dose reconstruction system, synchronized with the treatment delivery and motion of the patient, which can provide the necessary feedback on the quality of the delivery.<i>Approach</i>. A GPU-based analytical dose engine capable of millisecond dose calculation for carbon ion therapy has been developed and interfaced with the next generation of the dose delivery system (DDS) in use at Centro Nazionale di Adroterapia Oncologica (CNAO). The system receives the spot parameters and the motion information of the patient during the treatment and performs the reconstruction of the planned and delivered 4D-doses. After each iso-energy layer, the results are displayed on a graphical user interface by the end of the spill pause of the synchrotron, permitting verification against the reference dose. The framework has been verified experimentally at CNAO for a lung cancer case based on a virtual phantom 4DCT. The patient's motion was mimicked by a moving Ionization Chamber (IC) 2D-array.<i>Main</i><i>results</i>. For the investigated static and 4D-optimized treatment delivery cases, real-time dose reconstruction was achieved with an average pencil beam dose calculation speed up to more than one order of magnitude smaller than the spot delivery. The reconstructed doses have been benchmarked against offline log-file based dose reconstruction with the TRiP98 treatment planning system, as well as QA measurements with the IC 2D-array, where an average gamma-index passing rate (3%/3 mm) of 99.8% and 98.3%, respectively, were achieved.<i>Significance</i>. This work provides the first real-time 4D-dose reconstruction engine for carbon ion therapy. The framework integration with the CNAO DDS paves the way for a swift transition to the clinics.

Physics in Medicine and BiologyVol. 69(20)
Istituto Nazionale di Fisica Nucleare (IT), GSI Helmholtz Centre for Heavy Ion Research (DE), Universidade Estadual de Santa Cruz (BR), Technische Universität Darmstadt (DE), Istituto Nazionale di Fisica Nucleare, Sezione di Torino (IT), National Center for Oncological Hadrontherapy (IT), University of Turin (IT), University of Naples Federico II (IT)
Openalex Percentile: Top 12%
Radiation Therapy and Dosimetry
2.09
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.