Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range

This study reports the effects of changes in the waveform and frequency of radio frequency (RF) energy on the tissue ablation range. We developed a 70-watt RFA generator that provides sine and square waves and allows frequency control between 10 Hz and 500 kHz. The changes in the ablation range according to the waveform and frequency were observed using the developed generator. In the waveform variation test, the distance between the electrodes and the electrode type were changed for both waveforms with the frequency set to 500 kHz. In the frequency variation test, the waveform and electrode type were changed with the frequency set to 10, 100, and 500 kHz, while the distance between the electrodes was set to 20 mm. A fixed 45 voltage was applied using the bipolar method. RF energy was applied for 90 s in vitro. The temperature was regulated to not exceed 70°C. The ablation range was calculated using ImageJ software. The analysis results showed that the ablation range was larger with the square wave than with the sine wave and at 10 kHz than at 500 kHz. The developed generator can advance research on ablation area and depth in RF ablation.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2024-09-19
DOI
https://doi.org/10.1371/journal.pone.0308691
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range

Jung‐Hoon Park, Jinsu An, Ki-Hyeon Park, Dong‐Sung Won et al.
PLoS ONE
Advanced MRI Techniques and Applications
article

Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range

Jung‐Hoon Park, Jinsu An, Ki-Hyeon Park, Dong‐Sung Won, Yubeen Park, Hyung-Sik Kim, Jiho Lee
article en

Abstract

This study reports the effects of changes in the waveform and frequency of radio frequency (RF) energy on the tissue ablation range. We developed a 70-watt RFA generator that provides sine and square waves and allows frequency control between 10 Hz and 500 kHz. The changes in the ablation range according to the waveform and frequency were observed using the developed generator. In the waveform variation test, the distance between the electrodes and the electrode type were changed for both waveforms with the frequency set to 500 kHz. In the frequency variation test, the waveform and electrode type were changed with the frequency set to 10, 100, and 500 kHz, while the distance between the electrodes was set to 20 mm. A fixed 45 voltage was applied using the bipolar method. RF energy was applied for 90 s in vitro. The temperature was regulated to not exceed 70°C. The ablation range was calculated using ImageJ software. The analysis results showed that the ablation range was larger with the square wave than with the sine wave and at 10 kHz than at 500 kHz. The developed generator can advance research on ablation area and depth in RF ablation.

PLoS ONEVol. 19(9)
Ulsan College (KR), Konkuk University (KR), Asan Medical Center (KR), University of Ulsan (KR)
Korea Medical Device Development Fund, Ministry of Trade, Industry and Energy, Ministry of Food and Drug Safety, Ministry of Science and ICT, South Korea
Affordable and clean energy
Openalex Percentile: Top 74%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range — Jung‐Hoon Park, Jinsu An, et al. · PLoS ONE (2024) | TGRS Research Map | TGRS