Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation
In PCV mode this ventilator provided relatively stable VT across clinically relevant PI values to ambient pressures as high as 284 kPa. However, because Fpeak decreases at higher ambient pressure, some user adjustment might be necessary for precise VT maintenance during clinical use at higher PIs and ambient pressures.