Design and performance of indium seals for size-constrained tunable laser spectrometers

Indium seals have been used extensively in ultra-high vacuum and cryogenic applications. Typically, these seals use indium alongside or in place of other metal gaskets in stainless-steel vacuum flanges, with some custom applications for flanges sealing directly with glass (optics or tubes). Here, we present the design and performance of three pressed indium seals (99.99% In) between aluminum and 0.5 in. diameter sapphire optics and aluminum and gold coated Kovar semiconductor packages (TO-66 and TO-39). Test fixtures were designed to mimic those of future tunable diode laser spectrometers for Earth, planetary, and manned spaceflight environmental monitoring applications. Successful high-hermeticity seals [<10-10 atm cc/s (He)] were achieved for all seals formed with sufficient pressure applied to allow indium to flow between mating surfaces. The hermeticity of the seals was maintained after temperature cycling (-10 to +80 °C, 20 cycles), with the optical seals surviving extended duration tests (-55 to +85 °C, per MIL-STD-883). Semiconductor packages (TO-39) subjected to these extended tests saw a moderate increase in leak rate [∼5 × 10-9 atm cc/s (He)]; however, further testing showed that either the glass-metal package seals or the indium were affected (the sample size was too small to draw firm conclusions for future applications). Overall, these results suggest long-term survivability of indium seals for Kovar-aluminum and sapphire-aluminum interfaces [>10 years at 10-10 atm cc/s (He)], where the coefficient of thermal expansion differs by approximately four times.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2024-09-01
DOI
https://doi.org/10.1063/5.0206815
Citations
1
Primary Topic
Spectroscopy and Laser Applications
Type
article
Field-Weighted Citation Impact
0.24

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and performance of indium seals for size-constrained tunable laser spectrometers

Rafal Krzysiak, L. E. Christensen, Mathieu Fradet, J. Kent Wallace et al.
1 citations
Review of Scientific Instruments
Spectroscopy and Laser Applications
0.24
article

Design and performance of indium seals for size-constrained tunable laser spectrometers

Rafal Krzysiak, L. E. Christensen, Mathieu Fradet, J. Kent Wallace, Frank A. F. Winiberg, Aaron C. Noell, YangQuan Chen, Kamjou Mansour
article en
1 citations

Abstract

Indium seals have been used extensively in ultra-high vacuum and cryogenic applications. Typically, these seals use indium alongside or in place of other metal gaskets in stainless-steel vacuum flanges, with some custom applications for flanges sealing directly with glass (optics or tubes). Here, we present the design and performance of three pressed indium seals (99.99% In) between aluminum and 0.5 in. diameter sapphire optics and aluminum and gold coated Kovar semiconductor packages (TO-66 and TO-39). Test fixtures were designed to mimic those of future tunable diode laser spectrometers for Earth, planetary, and manned spaceflight environmental monitoring applications. Successful high-hermeticity seals [<10-10 atm cc/s (He)] were achieved for all seals formed with sufficient pressure applied to allow indium to flow between mating surfaces. The hermeticity of the seals was maintained after temperature cycling (-10 to +80 °C, 20 cycles), with the optical seals surviving extended duration tests (-55 to +85 °C, per MIL-STD-883). Semiconductor packages (TO-39) subjected to these extended tests saw a moderate increase in leak rate [∼5 × 10-9 atm cc/s (He)]; however, further testing showed that either the glass-metal package seals or the indium were affected (the sample size was too small to draw firm conclusions for future applications). Overall, these results suggest long-term survivability of indium seals for Kovar-aluminum and sapphire-aluminum interfaces [>10 years at 10-10 atm cc/s (He)], where the coefficient of thermal expansion differs by approximately four times.

Review of Scientific InstrumentsVol. 95(9)
Jet Propulsion Laboratory (US), University of California, Merced (US)
Advanced Exploration Systems
Openalex Percentile: Top 50%
Spectroscopy and Laser Applications
0.24
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.