Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited)

As part of a program to measure phase transition timescales in materials under dynamic compression, we have designed new x-ray imaging diagnostics to record multiple x-ray diffraction measurements during a single laser-driven experiment. Our design places several ns-gated hybrid CMOS (hCMOS) sensors within a few cm of a laser-driven target. The sensors must be protected from an extremely harsh environment, including debris, electromagnetic pulses, and unconverted laser light. Another key challenge is reducing the x-ray background relative to the faint diffraction signal. Building on the success of our predecessor (Target Diffraction In Situ), we implemented a staged approach to platform development. First, we built a demonstration diagnostic (Gated Diffraction Development Diagnostic) with two hCMOS sensors to confirm we could adequately protect them from the harsh environment and also acquire acceptable diffraction data. This allowed the team to quickly assess the risks and address the most significant challenges. We also collected scientifically useful data during development. Leveraging what we learned, we recently developed a much more ambitious instrument (Flexible Imaging Diffraction Diagnostic for Laser Experiments) that can field up to eight hCMOS sensors in a flexible geometry and participate in back-to-back shots at the National Ignition Facility (NIF). The design also allows for future iterations, such as faster hCMOS sensors and an embedded x-ray streak camera. The enhanced capabilities of the new instrument required a much more complex design, and the unexpected issues encountered on the first few shots at NIF remind us that complexity has consequences. Our progress in addressing these challenges is described herein, as is our current focus on improving data quality by reducing x-ray background and quantifying the uncertainties of our diffraction measurements.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2024-09-01
DOI
https://doi.org/10.1063/5.0219574
Citations
2
Primary Topic
Laser-Plasma Interactions and Diagnostics
Type
article
Field-Weighted Citation Impact
1.32

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited)

Cara E. Vennari, A. J. Mackinnon, L. R. Benedetti, S. R. Nagel et al.
2 citations
Review of Scientific Instruments
Laser-Plasma Interactions and Diagnostics
1.32
article

Developing time-resolved x-ray diffraction diagnostics at the National Ignition Facility (invited)

Cara E. Vennari, A. J. Mackinnon, L. R. Benedetti, S. R. Nagel, Nathan E. Palmer, J. H. Eggert, D. K. Bradley, Y. Ping, N. Bhandarkar, Robert Petre, Peter R. Nyholm, A. C. Carpenter
article en
2 citations

Abstract

As part of a program to measure phase transition timescales in materials under dynamic compression, we have designed new x-ray imaging diagnostics to record multiple x-ray diffraction measurements during a single laser-driven experiment. Our design places several ns-gated hybrid CMOS (hCMOS) sensors within a few cm of a laser-driven target. The sensors must be protected from an extremely harsh environment, including debris, electromagnetic pulses, and unconverted laser light. Another key challenge is reducing the x-ray background relative to the faint diffraction signal. Building on the success of our predecessor (Target Diffraction In Situ), we implemented a staged approach to platform development. First, we built a demonstration diagnostic (Gated Diffraction Development Diagnostic) with two hCMOS sensors to confirm we could adequately protect them from the harsh environment and also acquire acceptable diffraction data. This allowed the team to quickly assess the risks and address the most significant challenges. We also collected scientifically useful data during development. Leveraging what we learned, we recently developed a much more ambitious instrument (Flexible Imaging Diffraction Diagnostic for Laser Experiments) that can field up to eight hCMOS sensors in a flexible geometry and participate in back-to-back shots at the National Ignition Facility (NIF). The design also allows for future iterations, such as faster hCMOS sensors and an embedded x-ray streak camera. The enhanced capabilities of the new instrument required a much more complex design, and the unexpected issues encountered on the first few shots at NIF remind us that complexity has consequences. Our progress in addressing these challenges is described herein, as is our current focus on improving data quality by reducing x-ray background and quantifying the uncertainties of our diffraction measurements.

Review of Scientific InstrumentsVol. 95(9)
Lawrence Livermore National Laboratory (US)
U.S. Department of Energy
Industry, innovation and infrastructure
Openalex Percentile: Top 11%
Laser-Plasma Interactions and Diagnostics
1.32
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.