Leg occlusion of the Ovation Alto stent graft main body caused by twisting of the stent graft: A case report

The Ovation Alto stent graft is a unique endovascular abdominal aortic aneurysm repair device. A low-viscosity, radiopaque liquid polymer fills the rings located within the legs of the device. We reported a rare case of a patient diagnosed with an abdominal aortic aneurysm who was surgically treated using endovascular abdominal aortic aneurysm repair with an Ovation Alto stent graft. Polymer underfilling due to contralateral leg twisting of the stent graft's main body resulted in limb occlusion. We highlighted the importance of avoiding the rotation of the Ovation Alto stent graft, thus preventing the potential twisting of this device within the sheath after the insertion of the stent graft's main body.

Authors

Publication Details

Journal
Journal of Vascular Surgery Cases and Innovative Techniques
Published
2024-09-17
DOI
https://doi.org/10.1016/j.jvscit.2024.101625
Primary Topic
Aortic aneurysm repair treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Leg occlusion of the Ovation Alto stent graft main body caused by twisting of the stent graft: A case report

Hayate Nomura, Yukihiro Hayatsu, Ryoichi Tsuruhara, M. Naganuma et al.
Journal of Vascular Surgery Cases and Innovative Techniques
Aortic aneurysm repair treatments
article

Leg occlusion of the Ovation Alto stent graft main body caused by twisting of the stent graft: A case report

Hayate Nomura, Yukihiro Hayatsu, Ryoichi Tsuruhara, M. Naganuma, Masaki Hata, Naoya Terao, Kazuhiro Yamaya
article en

Abstract

The Ovation Alto stent graft is a unique endovascular abdominal aortic aneurysm repair device. A low-viscosity, radiopaque liquid polymer fills the rings located within the legs of the device. We reported a rare case of a patient diagnosed with an abdominal aortic aneurysm who was surgically treated using endovascular abdominal aortic aneurysm repair with an Ovation Alto stent graft. Polymer underfilling due to contralateral leg twisting of the stent graft's main body resulted in limb occlusion. We highlighted the importance of avoiding the rotation of the Ovation Alto stent graft, thus preventing the potential twisting of this device within the sheath after the insertion of the stent graft's main body.

Journal of Vascular Surgery Cases and Innovative TechniquesVol. 10(6)
Openalex Percentile: Top 72%
Aortic aneurysm repair treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.