Wireless Ultrasound Probes: A New Frontier In Assessing Femoral Cartilage Health

Ultrasound (US) is a non-invasive imaging technique that can accurately assess cartilage health. Recent advancements in wireless US technology allows for point of care cartilage imaging, but validation against diagnostic units remains to be established for knee cartilage outcomes. PURPOSE: Estimate agreement of articular cartilage thickness and echo-intensity between standard and wireless US units. METHODS: Using standard and wireless US, femoral cartilage was assessed via transverse suprapatellar scans with the knee in maximum flexion in 71 female NCAA Division 1 athletes (age: 20.0 ± 1.3 years, height: 171.7 ± 8.7 cm, mass: 69.4 ± 11.0 kg). Images were captured successively in the same location removing the unit after each acquisition. The wireless US unit used an auto-gain, while the standard unit used a fixed gain. Each image was greyscale-normalized by rescaling to the most hyper- and hypoechoic pixels to account for unit settings. The global femoral cartilage was manually segmented in each image. A semi-automated program was used to quantify the mean thickness and echo-intensity of raw and normalized images for the global region of interest, as well as for the medial, middle, and lateral subregions. Intraclass correlation coefficients (ICC2,k) for absolute agreement, standard error of the measurement, and minimum detectable difference were calculated between the standard and wireless US units. RESULTS: Cartilage thickness demonstrated good to excellent agreement for all regions (ICC2,k = 0.88-0.96; Table 1). However, mean echo-intensity demonstrated poor agreement in all regions for raw (ICC2,k = 0.43-0.68) and normalized images (ICC2,k = 0.45-0.58). CONCLUSIONS: Wireless US units produce similar results for knee cartilage thickness compared to their diagnostic counterparts. However, further research is needed to understand the error involved when measuring echo-intensity using different US machines so that comparison across units is possible. Dr. Harkey was supported by a National Institute of Arthritis and Musculoskeletal and Skin Diseases (NIAMS) grant (K01 AR081389). ChatGPT was used to revise the content of this research abstract

Authors

Institutions

Publication Details

Journal
Medicine & Science in Sports & Exercise
Published
2024-09-16
DOI
https://doi.org/10.1249/01.mss.0001053124.19355.22
Primary Topic
Osteoarthritis Treatment and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Wireless Ultrasound Probes: A New Frontier In Assessing Femoral Cartilage Health

Ilker Hacihaliloglu, Corey D. Grozier, Brad Winn, Matthew S. Harkey et al.
Medicine & Science in Sports & Exercise
Osteoarthritis Treatment and Mechanisms
article

Wireless Ultrasound Probes: A New Frontier In Assessing Femoral Cartilage Health

Ilker Hacihaliloglu, Corey D. Grozier, Brad Winn, Matthew S. Harkey, Ryan Fajardo, Robert Dima, Jessica Tolzman, Arjun Parmar
article en

Abstract

Ultrasound (US) is a non-invasive imaging technique that can accurately assess cartilage health. Recent advancements in wireless US technology allows for point of care cartilage imaging, but validation against diagnostic units remains to be established for knee cartilage outcomes. PURPOSE: Estimate agreement of articular cartilage thickness and echo-intensity between standard and wireless US units. METHODS: Using standard and wireless US, femoral cartilage was assessed via transverse suprapatellar scans with the knee in maximum flexion in 71 female NCAA Division 1 athletes (age: 20.0 ± 1.3 years, height: 171.7 ± 8.7 cm, mass: 69.4 ± 11.0 kg). Images were captured successively in the same location removing the unit after each acquisition. The wireless US unit used an auto-gain, while the standard unit used a fixed gain. Each image was greyscale-normalized by rescaling to the most hyper- and hypoechoic pixels to account for unit settings. The global femoral cartilage was manually segmented in each image. A semi-automated program was used to quantify the mean thickness and echo-intensity of raw and normalized images for the global region of interest, as well as for the medial, middle, and lateral subregions. Intraclass correlation coefficients (ICC2,k) for absolute agreement, standard error of the measurement, and minimum detectable difference were calculated between the standard and wireless US units. RESULTS: Cartilage thickness demonstrated good to excellent agreement for all regions (ICC2,k = 0.88-0.96; Table 1). However, mean echo-intensity demonstrated poor agreement in all regions for raw (ICC2,k = 0.43-0.68) and normalized images (ICC2,k = 0.45-0.58). CONCLUSIONS: Wireless US units produce similar results for knee cartilage thickness compared to their diagnostic counterparts. However, further research is needed to understand the error involved when measuring echo-intensity using different US machines so that comparison across units is possible. Dr. Harkey was supported by a National Institute of Arthritis and Musculoskeletal and Skin Diseases (NIAMS) grant (K01 AR081389). ChatGPT was used to revise the content of this research abstract

Medicine & Science in Sports & ExerciseVol. 56(10S)
Radiology Associates (US), Michigan State University (US)
Zero hunger
Openalex Percentile: Top 72%
Osteoarthritis Treatment and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.